1.
With a specimen mover plate mounted, place a
preparation surface on the preparation disc.
2.
Select the ‘thickest’ preparation surface to be used and
place it on the preparation disc. Usually, this will be SiC
Foil on an MD-Gekko disc, or SiC Paper on an MD-
Fuga disc, or an MD-Alto.
3.
Place the supplied spacing disc on the preparation
surface.
A
Spacing disc
4.
Support the LaboForce-100 head and loosen the
screw in the adjustment ring.
A
V-shaped groove
B
Pin
C
Adjustment ring
D
Fastening screw
5.
Press the
Lower/Raise
button to lower the specimen
mover head. An error message will appear as the
specimen mover head is not in contact with the
adjustment ring.
5 Installation
LaboForce-100
13