In the case of electrical products, individual components or the product itself may be live. Live
parts can cause electric shocks in the event of contact and injure patients, users, and third
parties.
Do not open the product.
Have servicing carried out by or a company authorized by . Failure to observe this will void
the guarantee.
Do not touch the output jacks of the product and the patient at the same time during use.
Always pull out the mains plug before carrying out any cleaning and maintenance work.
Disconnect the power plug before opening the housing.
3.7 Damage due to ingress of liquid in electrical
components
In the case of electrical products, individual components or the product itself may be live.
Liquid ingress into an electrical product may result in a short circuit or an unintentional transfer
of current. The product is damaged as a result and patients, users and third parties may be
injured.
Do not store liquids near the product or on the product.
If liquid has entered the product, pull out the plug and allow the product to dry completely.
3.8 Risk of explosion
The product can generate sparks, which cause combustible or flammable gases and liquids to
ignite or explode. This may cause injuries to patients, users, and third parties.
The product must not be operated in oxygenated environments.
Do not operate the product in explosive atmospheres.
Do not operate the product in environments with combustible gases such as inhalation
anesthetics and mixtures thereof. Observe the hazard zones:
Only connect or disconnect the power plug to or from the power supply outside explosive
atmospheres.
Combustible and flammable gases must be allowed to escape, be extracted, or be
displaced with CO2 before use.
Combustible and flammable liquids must be allowed to vaporize or be extracted before
use.
Only start the application when combustible or flammable gases and liquids are no longer
present.
3.9 Observing ambient conditions
If the product is used in an unsuitable environment, patients, users, and third parties may be
injured.