Electromagnetic compatibility
Instructions for use • C-MAC Monitor • NDQ566_EN_V1.0_02-2021_IFU_CE-MDR
43
Emission measurements
Compliance
Electromagnetic environment –
Guidelines
Voltage fluctuations/flicker emis-
sions as per IEC 61000-3-3
complies
power supply network that supplies
buildings used for domestic pur-
poses.
9.6 Table 5 – Recommended separation distances between
portable and mobile HF communications equipment
and the product
The product is intended for use in an electromagnetic environment in which HF disturbances
are controlled. The customer or user of the product can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile HF
communications equipment (transmitters) and the product as recommended below, according
to the output energy of the communications equipment.
Rated power of the
transmitter [W]
Separation distance d [m] according to the transmitter frequency
150 kHz to 80 MHz
d = 1.2 √P
80 MHz to 800 MHz
d = 1.2 √P
800 MHz to 2.5 GHz
d = 2.3 √P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters whose maximum rated power is not listed in the table above, the recom-
mended separation distance d in meters (m) can be estimated using the equation from the re-
spective column, whereby P is the maximum rated power of the transmitter in watts (W) ac-
cording to the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range ap-
plies.
Note: These guidelines may not apply in all situations. The propagation of electromagnetic
waves is affected by absorptions and reflections of buildings, objects, and people.