English
Page | 22
P
ATIENT
I
DENTIFICATION
C
ARD
A patient identification card (ID) is mailed to you after Stimwave receives your
registration form. The patient ID card supplies information about you, your
stimulator system, and your doctor. Your identification cad may allow you to
bypass security devices. Carry this card with you at all times. If you move,
change doctors, or lose your card, contact Stimwave for a replacement card.
Refer to the Stimwave contacts at the end of this manual.
S
PECIFICATIONS
Item
Specification
Amplitude
0 to 12.8 mA per electrode pair
Pulse Width
50 to 500 µs
Frequency (of therapy)
5 to 1,500 Hz
Number of Channels
2
Number of Programs (selectable)
3
Transmit Carrier Frequency
915 MHz
Bluetooth Frequency
2.402 GHz
Charger Power Source
120-240V 50-60 Hz power line
Operating/storage temperature,
relative humidity
-18
º
C to 60
º
C (0
º
F to 140
º
F)
20% to 90%
Operating/storage atmospheric
pressure
70 kPa to 150 kPa
(20.7 in Hg to 44.3 in Hg)
Size (approximate)
7.6 cm x 5 cm x 2 cm
(3 in x 2 in x 0.8 in)
Weight (approximate)
0.5 kg (1 lb.)
WAA Material (Do not place directly
on skin, to be worn over a thin layer
of clothing at all times)
Aluminum
Parameter
Min
Typical
Max
Units
Supply Voltage
120
---
240
V
A/C
Supply Voltage Frequency
50
---
60
Hz
Output Voltage (DC)
4.5
5
5.5
V
Input Current
---
---
1.0
A
Output Current (each)
---
---
2.4
A
Battery Recharge Time (assuming
3.7V battery)
2
3
4
hours
Modulation
---
CW
---
N/A