Instruction manual for the hospital bed Deka _____________________________________________
Page
15
3.1
C
HECKLIST
:
I
NSPECTION BY THE USER
Inspection
ok
Not
ok
Description of
defect
Visual inspection of the mechanical components
Patient’s lifting pole,
location sleeves
Damage, deformation
Grab handle with strap Damage
Undercarriage
Damage, deformation
Mattress base, covers
Damage, deformation
Performance check of the mechanical components (if installed)
Castors
Locking, moving
Adjusting the backrest Functional
test
according
to
instruction manual
Side guards
Functional
test,
securely engaged
Lower leg rest
Functional test
Accessories
(e.g.
patient’s lifting pole,
grab handle with strap,
bed rails)
Fastenings, damage;
suitability
Inspector’s signature:
Inspection results:
Date:
3.2
L
OCATION
R
EQUIREMENTS
There must be sufficient room available to accommodate the bed’s entire
range of adjustments. Obstacles such as bedside lockers, floor cable ducts,
furniture, or window sills must not impede adjustments.
Before using the bed on parquet flooring, check whether the castors can
leave marks on the parquet varnish. We accept no liability for such wear. The
bed can be used on tiles, carpet, linoleum or laminate flooring without causing
any damage.
If other equipment (e.g. compressors for positioning systems, etc.) is attached,
ensure that all items of equipment are fixed securely and function properly.
Pay special attention here to the safe routing of all loose connector cables,
tubing, etc. If you have any queries or concerns, consult the manufacturer of
the equipment or Stiegelmeyer.