LIMITED WARRANTY
Sterisil
®
warrants the Sterisil
®
System G5 to be free from defects in parts, materials, manufacturing, or labor
for a period of one year after the device is placed in service by the end user (the “Warranty Period”), and on
select electrical components from the date of Sterisil’s Purchase. This Limited Warranty does not include user
replaceable parts required for routine maintenance of the product, or damage resulting from abuse or neglect
of the product, and covers only use of the product for its intended purpose, and not for any other use. End
users should contact Sterisil or an authorized distributor for claims arising during the Warranty Period. Sterisil
reserves the right to repair or replace, in its discretion, any product that may exhibit a defect during the Warranty
Period. Please note that the bladder tanks need to be replaced approximately every 2 years.
EXCEPT FOR THE FOREGOING LIMITED WARRANTY, ALL Sterisil
®
PRODUCTS ARE SOLD “AS IS.” THE
FOREGOING LIMITED WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR
IMPLIED, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
Sterisil
®
shall not be liable to any user of any product, or any other person, for any indirect, special, incidental,
exemplary or consequential damages (including, without limitation, lost profits) resulting from use or inability to
use any Sterisil product, arising from any cause of action whatsoever, including contract, warranty, strict liability,
or negligence, even if Sterisil has been notified of the possibility of such damages. Under no circumstances shall
the liability of Sterisil to the user or any other party exceed the purchase price of the product. No action under
the foregoing limited warranty may be brought more than one (1) year after the expiration of the Warranty Period.
Exclusive subject matter and personal jurisdiction for all disputes arising under this Limited Warranty shall be in
the El Paso County District Court in Colorado Springs, Colorado.
STERISIL
®
RECYCLING PROGRAM
Sterisil offers a free recycling program, allowing customers to recycle their used Sterisil
®
Cartidges by a click of a button. Simply visit
sterisil.com/green
or call us directly to
request a free shipping label.
While other companies dispose
100%
of their used products to a landfill, Sterisil is able to
recycle
91%
of our Straws and Cartridges. In addition to recycling, Sterisil contributes to a
greener environment by being the only dental waterline company to offer a residual
disinfectant that meets BMP compliance by not increasing the mercury output in
amalgam separators.
1. Start by unplugging the system from the power outlet.
2. Locate the UV housing behind the cartridges. Pull down on the UV power cable to feed slack for removing the
bulb.
3. Press in on UV lamp harness (not the ribbed housing sleeve) on the right side of the housing and twist
counterclockwise until the bulb releases.
4. Pull the bulb completely out of the housing.
5. Separate the bulb from the cable.
6. Remove the protective wrapping from the new bulb (grab bulb by green ceramic sections only or wear clean
rubber gloves as a precaution).
7. Connect the bulb to the cable.
8. Slide bulb all the way in and carefully rotate clockwise until harness seats itself.
9. Push the excess length of cable back into the system.
10. Plug In system and wait for full power up.
11. Follow the procedure from Phase 5 in the installation manual to reset the monitor.
CUSTOMER SERVICE
Phone number: 719 622 7200 then press the option for “Customer Service” when prompted.
Cartridges
Lifespan
Stage 1
1 year
Stage 2
9,463L
Stage 3
1 year or the TDS water quality
Stage 4A
Dependent on water quality and usage
Stage 4B
Dependent on water quality and usage
Class A UV Light
Dependent on water quality and usage
Stage 5 - 1,000L
1,000 liters of dental water
Stage 5 - 3,000L
3,000 liters of dental water
Stage 5 - 7,000L
7,000 liters of dental water
Stage 5 - 10,000L
10,000 liters of dental water
CARTRIDGE LIFE TABLE
CHANGING THE CLASS A UV LIGHT
The Class A UV light on the Sterisil
®
System G5 is a powerful tool in bacteria reduction. Do not remove the UV
bulb with the system plugged in. Never look directly at the bulb when it is powered up and emitting light. Removal
and replacement of an expired bulb takes less than 5 minutes and requires no special tools. Replacement may be
easier with the cartridges removed from port 5 and 6. Always wear clean gloves when handling a new UV bulb.