MAINTENANCE MANUAL

Thanks to an integrated network of sensors and smart technology, the Sterisil

®

 System G5 is easier than ever to 

maintain. Visual alerts and audible alarms display prominently on the touchscreen and the PC app the issue at hand.

ALERTS - ATTENTION REQUIRED

A yellow bar will appear in the top of the touchscreen 

or on the PC app when a system function nears the 

threshold of normal operating parameters. Simply touch 

the yellow bar to reveal the details, and follow the “next 

step alert guidance” system to resolve the issue.  

ALARMS - REQUIRES IMMEDIATE SERVICE

An alarm with a red bar indicates an issue demanding 

immediate attention. “Alarm: System Requires 

Immediate Service.” Simply touch the red bar or click 

on the PC app to reveal the details, and follow the “next 

step alert guidance” system to resolve the issue.  

You may also bring up the details view by touching 

either the desired Stage number from the “Home” 

screen or select the “Stages” tab from the bottom row. 

STEP BY STEP FILTER REPLACEMENT 

INSTRUCTIONS:

Thanks to internal valving, cartridge change outs 

are simple and easy. It is not necessary to shut off 

the source water from the supply tee. Water flow is 

automatically shut off when a cartridge is removed. 

Please have a towel nearby for excess water in 

cartridges.

1.  Before removing a cartridge, place a large “X” on 

the one to be replaced to avoid any confusion when the new 

cartridge is introduced.

2.  If you are changing the Stage 3 Cartridge, it will be necessary to 

remove the brine tube from the bottom of the cartridge prior to 

removal. Simply push in on the collet and pull out on the tubing 

to remove. You will reinstall this tube by pressing the tubing into 

the collet once the new Stage 3 Cartridge is in place. 

3.  To remove a cartridge, rotate counter clockwise until the 

cartridge is free of the valve head. A small amount of water may 

escape during this process. Don’t panic, this is normal. 

4.  Before placing a new cartridge in the system, write the date in 

the space provided on the label. Remove the yellow cap from 

the top of cartridge, and rotate the cartridge clockwise into the 

valve head to secure it. Ensure cartridge is fully screwed in.

5.  Reprogram the capacity monitoring from the touchscreen as referenced in Phase 5 of the installation 

procedure. 

6.  The final step when changing cartridges is to run some water for 60 seconds. For change outs on Stages 1-4, 

close the RO tank and run the autoclave water. For Stage 5 change outs, close the Dental Storage Tank and 

run the dental water (Stage 5). 

7.  Open all tanks to finish.

8.  Cartridges can either be discarded in the trash or returned to Sterisil for recycling. For more information on 

our recycling program visit sterisil.com/recycle.

• Make a clean square cut on the source line using a tube cutter or sharp knife. This cut should be close to 

the Pre-filter Kit if included. There are arrows on the Pump to show the direction of water flow. Insert the end 

coming from the municipal water source into the intake of the Booster Pump. The other end of the tubing will 

go into the outlet on the Booster Pump. Push tubing until it is fully inserted into each fitting. 

• On the RO tank line, make a clean, square cut as close to the tank as possible. Your Pressure Switch will 

fit into this line. Push tubing until it is fully inserted into the fittings on the Pressure Switch. This Switch is 

multidirectional.

• The Pressure Switch will connect to both the white ballast and the power cord on the Booster Pump. 

• Turn your source water back on, plug in the power cord, and run autoclave water for 60 seconds. 

• Check for any leaks. If there are any leaks, ensure pieces are properly connected. 

STERISIL

®

 WAND (OPTIONAL COMPONENT)

• Use a drill to create a hole in which to put the anchor using a 15/64” to ¼” drill bit. 

• Screw the #8 screws into the anchor through the wand holder. 

• Unscrew the first cartridge to stop the water to your system. You may need a towel to catch water that may 

drip. 

• Cut the tubing coming out of the autoclave port on the system approximately 6-8 inches from the system. Use 

shard scissors or knife to cut tubing.

• Connect the Union T to this junction. Push until tubing is fully inserted into the fitting on each side. 

• With the remaining female end of the Union T, connect the blue coiled hose with the inline ball valve opposite 

the Union T.

• Attach the Wand to the blue ¼” tubing. 

• Screw the first cartridge in to start water flow. 

• Check for any leaks. If there are any leaks, make sure that the pieces are completely connected. 

• The head of the Wand rotates to change the function. For small necked bottles, turn the head counterclockwise 

to make the water come out in a single stream. For large neck bottles and autoclaves, a multi stream function 

can be activated by turning the head of the wand clockwise. 

• IMPORTANT- Make sure to turn off the ball valve between uses of the Wand to prevent leakage.

®

PLAN NOTES

18

11

9

13

7

4

5

10

16

14

14

Summary of Contents for System G5

Page 1: ...INSTALLATION AND MAINTENANCE MANUAL GP9003 A ...

Page 2: ...the user to see real time data from a network of on board sensors for convenient system health monitoring In the event of a maintenance alert users can access built in troubleshooting guides pertaining to their specific error for quick and easy resolution 835 S Highway 105 Suite D Palmer Lake CO 80133 Office 719 622 7200 Fax 866 299 2495 Sterisil offers optional add ons to the G5 to ensure proper ...

Page 3: ...r go to the settings tab on the touchscreen Record the installation date on the warranty card Please ensure you send in the warranty card or register online The initial set up can take up to 3 5 hours depending on customer specifications Ensure the lengths of tubing provided will meet the needs of your specific configuration All tubing cuts should be made at a right angle to the tubing itself Be s...

Page 4: ...of mounting the system Sterisil recommends anchoring the system into a wall stud The EZ toggle anchors should only be used when drilling into drywall STATIC MOUNT BAG B 1 Affixing the Static Plate to the Wall Identify the slotted holes on the plate These slots will be used to mount the plate to the wall The static plate is symmetrical so any side can be up Mock up your static mounting plate in the...

Page 5: ...ERIPHERAL COMPONENTS Peripheral components No more than 50 feet from the system Depending on the customer s specific configuration you may have optional components such as the Pre filter Kit and Booster Pump plumbed between the municipal source and the red SOURCE port on the system G5 POWER ADAPTER Select location within 6 feet of the outlet Ensure there is no tension on the cable If you choose to...

Page 6: ... D Tank Port Blue fitting Indicator Run 1 4 tubing from the blue DENTAL port on the system to the 1 4 Tank Ball Valve on the smaller dental water storage tank Autoclave Port Green fitting indicator OPTIONAL ADD ON Sterisil Wand see page 14 Run 1 4 tubing from the green AUTOCLAVE port on the system to the white Autoclave Faucet Drain Port Black fitting indicator Bag D Run a 1 4 line from the black ...

Page 7: ...DD ONS PRE FILTER KIT OPTIONAL COMPONENT Considerations Account for a minimum of 2 clearance at the bottom for future filter replacements Location of Pre filter and Inline Ball Valve diagram is on page 5 Turn off your source water Find a location where you can drive the screws for the bracket into the wall that will reach both your source water line and your Sterisil System G5 source line If you h...

Page 8: ...for 60 seconds For change outs on Stages 1 4 close the RO tank and run the autoclave water For Stage 5 change outs close the Dental Storage Tank and run the dental water Stage 5 7 Open all tanks to finish 8 Cartridges can either be discarded in the trash or returned to Sterisil for recycling For more information on our recycling program visit sterisil com recycle Make a clean square cut on the sou...

Page 9: ...ed Sterisil Cartidges by a click of a button Simply visit sterisil com green or call us directly to request a free shipping label While other companies dispose 100 of their used products to a landfill Sterisil is able to recycle 91 of our Straws and Cartridges In addition to recycling Sterisil contributes to a greener environment by being the only dental waterline company to offer a residual disin...

Reviews: