MAINTENANCE MANUAL
Thanks to an integrated network of sensors and smart technology, the Sterisil
®
System G5 is easier than ever to
maintain. Visual alerts and audible alarms display prominently on the touchscreen and the PC app the issue at hand.
ALERTS - ATTENTION REQUIRED
A yellow bar will appear in the top of the touchscreen
or on the PC app when a system function nears the
threshold of normal operating parameters. Simply touch
the yellow bar to reveal the details, and follow the “next
step alert guidance” system to resolve the issue.
ALARMS - REQUIRES IMMEDIATE SERVICE
An alarm with a red bar indicates an issue demanding
immediate attention. “Alarm: System Requires
Immediate Service.” Simply touch the red bar or click
on the PC app to reveal the details, and follow the “next
step alert guidance” system to resolve the issue.
You may also bring up the details view by touching
either the desired Stage number from the “Home”
screen or select the “Stages” tab from the bottom row.
STEP BY STEP FILTER REPLACEMENT
INSTRUCTIONS:
Thanks to internal valving, cartridge change outs
are simple and easy. It is not necessary to shut off
the source water from the supply tee. Water flow is
automatically shut off when a cartridge is removed.
Please have a towel nearby for excess water in
cartridges.
1. Before removing a cartridge, place a large “X” on
the one to be replaced to avoid any confusion when the new
cartridge is introduced.
2. If you are changing the Stage 3 Cartridge, it will be necessary to
remove the brine tube from the bottom of the cartridge prior to
removal. Simply push in on the collet and pull out on the tubing
to remove. You will reinstall this tube by pressing the tubing into
the collet once the new Stage 3 Cartridge is in place.
3. To remove a cartridge, rotate counter clockwise until the
cartridge is free of the valve head. A small amount of water may
escape during this process. Don’t panic, this is normal.
4. Before placing a new cartridge in the system, write the date in
the space provided on the label. Remove the yellow cap from
the top of cartridge, and rotate the cartridge clockwise into the
valve head to secure it. Ensure cartridge is fully screwed in.
5. Reprogram the capacity monitoring from the touchscreen as referenced in Phase 5 of the installation
procedure.
6. The final step when changing cartridges is to run some water for 60 seconds. For change outs on Stages 1-4,
close the RO tank and run the autoclave water. For Stage 5 change outs, close the Dental Storage Tank and
run the dental water (Stage 5).
7. Open all tanks to finish.
8. Cartridges can either be discarded in the trash or returned to Sterisil for recycling. For more information on
our recycling program visit sterisil.com/recycle.
• Make a clean square cut on the source line using a tube cutter or sharp knife. This cut should be close to
the Pre-filter Kit if included. There are arrows on the Pump to show the direction of water flow. Insert the end
coming from the municipal water source into the intake of the Booster Pump. The other end of the tubing will
go into the outlet on the Booster Pump. Push tubing until it is fully inserted into each fitting.
• On the RO tank line, make a clean, square cut as close to the tank as possible. Your Pressure Switch will
fit into this line. Push tubing until it is fully inserted into the fittings on the Pressure Switch. This Switch is
multidirectional.
• The Pressure Switch will connect to both the white ballast and the power cord on the Booster Pump.
• Turn your source water back on, plug in the power cord, and run autoclave water for 60 seconds.
• Check for any leaks. If there are any leaks, ensure pieces are properly connected.
STERISIL
®
WAND (OPTIONAL COMPONENT)
• Use a drill to create a hole in which to put the anchor using a 15/64” to ¼” drill bit.
• Screw the #8 screws into the anchor through the wand holder.
• Unscrew the first cartridge to stop the water to your system. You may need a towel to catch water that may
drip.
• Cut the tubing coming out of the autoclave port on the system approximately 6-8 inches from the system. Use
shard scissors or knife to cut tubing.
• Connect the Union T to this junction. Push until tubing is fully inserted into the fitting on each side.
• With the remaining female end of the Union T, connect the blue coiled hose with the inline ball valve opposite
the Union T.
• Attach the Wand to the blue ¼” tubing.
• Screw the first cartridge in to start water flow.
• Check for any leaks. If there are any leaks, make sure that the pieces are completely connected.
• The head of the Wand rotates to change the function. For small necked bottles, turn the head counterclockwise
to make the water come out in a single stream. For large neck bottles and autoclaves, a multi stream function
can be activated by turning the head of the wand clockwise.
• IMPORTANT- Make sure to turn off the ball valve between uses of the Wand to prevent leakage.
®