6-10
10087252
Operator ManualSupplemental Information
Figure 6-8. Component Identification
3.
Cup Draw Tube – Draws Sterilant from Sterilant Cup.
4.
Cup – Disposable Cup supplies Vaprox 59 Hydrogen Peroxide
Sterilant.
5.
USB Port – Biodecontamination Unit outputs cycle data in
encrypted format for data storage through this port.
6.
ON/OFF Switch – Enables power to controller.
7.
Start Touch Pad – When pressed, starts cycle.
8.
Abort Touch Pad - When pressed, aborts cycle.
9.
Operation Screen – Can contain up to four lines of Operational
text to enable operator to communicate with controller.
6.10 Operation
The following information is intended to enable a Trained and Certified
Applicator to properly use the Biodecontamination Unit to
Biodecontaminate an Enclosure (refer to
):
The STERIS X10 Biodecontamination Unit is to be used by Trained
and Certified Applicators who have successfully completed both the
STERIS Training and Certification Course for applicators of Vaprox 59
Hydrogen Peroxide Sterilant and the STERIS X10 Biodecontamination
Unit Operator Course. Certification must be active and in force for all
Applicators of Vaprox 59 Hydrogen Peroxide Sterilant.
1.
Ensure Biodecontamination Unit is properly attached to Enclosure
(refer to
).
To Sash Plate
Assembly
To Exhaust Plenum
Adapter
Supply Door
Fastening Screws
Desiccant
Cartridge
Cup Draw Tube and Lid,
and Sterilant Cup
ON/OFF Switch
Start Touch Pad
Abort Touch Pad
Operation Screen
Power Cord
RS-232 Port
USB Port
Summary of Contents for VHP X10
Page 1: ...OPERATOR MANUAL X10 Biodecontamination Unit TEAM 10087252 ...
Page 2: ......
Page 10: ...viii 129396 083 Operator Manual Table of Contents Life Sciences ...
Page 12: ...xii 10087252 Operator Manual Table of Contents Life Sciences ...
Page 20: ...2 4 10087252 Operator Manual Terms Definitions and Symbols Life Sciences ...