4-1
Troubleshooting
Operator Manual
10087252
4.1 Troubleshooting
The VHP
X10
™
Biodecontamination Unit alerts the Trained and
Certified Applicator of various alarm/abort situations. Alarms occur
as a result of component failures, utility failures, unexpected
measurements from sensors or as a warning indication that
Applicator intervention is necessary (see
Refer to Troubleshooting Guide (
) for description of Alarm
Message Displays.
If a situation occurs that is not described in this manual, please call
STERIS Life Sciences. Trained service personnel can promptly
restore this
X10
Biodecontamination Unit to proper working order.
NOTE: Never permit unqualified persons to service this
Biodecontamination Unit.
IMPORTANT: A listing of the
S
AFETY
P
RECAUTIONS
to be observed when operating and servicing this
equipment and/or handling Vaprox 59 Hydrogen Peroxide Sterilant can be found in
of this
manual. Do not operate or service the equipment or handle the Sterilant until you have been trained
and certified on this information as well as the information on the Vaprox 59 Hydrogen Peroxide
Sterilant product label and package insert.
DANGER – PERSONAL
INJURY, CONTAMINATED
ENCLOSURE AND/OR
EQUIPMENT DAMAGE
HAZARD: No one may enter a
sealed Enclosure during or
after an aborted cycle without
SCBA and appropriate PPE
(refer to Sterilant SDS for
PPE) if Sterilant levels within
the treated Enclosure are
above one PPM. Refer to
Sterilant label and package
insert for instructions.
WARNING – PERSONAL
INJURY AND/OR EQUIPMENT
DAMAGE HAZARD: Integrated
(temperature and humidity)
sensors are fragile. Do not
bang, twist or crush.
Biodecontamination Unit does
not operate properly with a
damaged sensor.
Table 4-1. Alarm/Abort Actions
Phase at Alarm-Abort
Action Following
Alarm-Abort
Before start of Sterilant
Injection phase (before
Condition/Biodecontamination)
Advance to Cycle Complete
During Sterilant injection
(Condition/Biodecontamination)
Advance to Aeration
Table 4-2. Power Interruption/Power-Up Actions
Active Cycle Phase
Action Upon Power
Return
Heat Cabinet
Abort Cycle
Dehumidify
Abort Cycle
Re-Heat Cabinet
Abort Cycle
Heater Cap Warm-Up
Abort Cycle
Condition
Aeration
Biodecontamination
Aeration
Aeration
Restart Aeration
Cycle Complete
Return to Cycle Complete
TROUBLESHOOTING
4
Summary of Contents for VHP X10
Page 1: ...OPERATOR MANUAL X10 Biodecontamination Unit TEAM 10087252 ...
Page 2: ......
Page 10: ...viii 129396 083 Operator Manual Table of Contents Life Sciences ...
Page 12: ...xii 10087252 Operator Manual Table of Contents Life Sciences ...
Page 20: ...2 4 10087252 Operator Manual Terms Definitions and Symbols Life Sciences ...