16. External Aerator Fail
Occurs when the External Aerator
option is ON and the External
Aerator input is open.
• Auxillary Aerator Unit unplugged from Unit:
Connect Auxillary Aerator Unit.
• External Aerator option inadvertently
selected: Deselect this option in the menu.
17. Sterilant Cup
Expired
Occurs during the reading of the
sterilant cup RFID tag, the sterilant
cup expiration date is past the
current calendar date
• Replace the sterilant cup with one that is
not expired.
• Check the sterilant cup
'
s printed expiration
date.
• Check that the BU system date is correct.
18. Sterilant Cup Used
Occurs during the reading of the
sterilant cup RFID tag, the sterilant
cup serial number matches a
previously used code
• Replace the sterilant cup with one that has
not been used.
19. Sterilant Cup Bad
SKU
Occurs during the reading of the
sterilant cup RFID tag, the sterilant
cup SKU code does not match the
BSC system SKU code
• Replace the sterilant cup with one that has
the correct SKU.
• Check the sterilant cup
'
s printed SKU code.
20. Sterilant Cup Bad
Wrt
Occurs during the condition
phase, the writing of the sterilant
cup RFID tag, and subsequent
verification reading, the RFID tag
'
s
serial number does not contain the
used cup code
• Manually mark the cup as used.
• Verify the proper operation of the RFID
Reader/Writer.
21. RH Sensor Not
Calib
Appears at any time the RH
Sensor
'
s Relative Humidity input is
identified as not calibrated
• System initiated: Not applicable.
22. RH RTD Not Calib
Appears at any time the RH
Sensor
'
s RTD input is identified as
not calibrated
• System initiated: Not applicable.
23. Machine Calib Due
Occurs any time calibration is due
for the BSC Unit analog inputs or
outputs
• System initiated: Not applicable.
24. Drum Assembly
Index Fail
Occurs at any time the drum
assembly does not encounter one
of the three indexer sensors
• Check sensors on the drum position board.
• Check the drum indexing motor (MTR-2).
• Verify proper torque on rear drum assembly
retaining bolt.
25. Control Board Fail
Occurs at any time a Control
Board fault is issued by the Control
Board
• TBD.
* STERIS Life Sciences
Table 4-3. Troubleshooting Guide (Cont’d)
Alarm Message
Display
Description
Possible Causes and / or
Corrections
Summary of Contents for VHP X10
Page 1: ...OPERATOR MANUAL X10 Biodecontamination Unit TEAM 10087252 ...
Page 2: ......
Page 10: ...viii 129396 083 Operator Manual Table of Contents Life Sciences ...
Page 12: ...xii 10087252 Operator Manual Table of Contents Life Sciences ...
Page 20: ...2 4 10087252 Operator Manual Terms Definitions and Symbols Life Sciences ...