7-21
764333-268
8. Remove six lower screws from table column plastic
cap (see Figure 7-25) and allow table column cover
to lower to base cover.
9. Remove seat section wire protection cover.
10. Remove right-side wire and hose protection cap
(only right side). See Figure 7-59.
Figure 7-59. Remove Back Sensor
11. Remove wiper from cylinder rod end (only right
side).
12. Disconnect sensor from CPU motherboard (P74
connector) and remove sensor.
13. Reassemble in reverse order and adjust tabletop
to level.
14. Verify cursor full movement is centered as much
as possible on sensing strip (does not approach
edges too closely).
7.35 HYDRAULIC FLUID
CONTAMINATION
This procedure is for use when hydraulic fluid
contamination is suspected inside the pump or one of
the manifolds.
1. Remove table column cover per
2. Remove appropriate base cover per
3. Unlock table and lower tabletop to lowest position.
4. Using a suction bottle and facility vacuum system,
remove all hydraulic oil from reservoir.
NOTE: Failure to use a suction bottle results in oil
contamination of facility vacuum system.
5. Refill reservoir to approximately 3/4 full or about
1" (25.4 mm) from top.
6. Fully articulate all table sections except slide five to
six times. Ensure floor locks and kidney elevator are
included.
7. Repeat
Steps 3 through 6
two more times.
8. Clean entire area outside all hose fittings to ensure
no additional dust or lint are introduced into
hydraulic system when disasssembled.
9.
Replace manifold(s) (per
) and/or pump and
motor assembly.
10. Remove block and tie-down.
11. Repeat
Steps 3 through 6
six more times.
12. Let table sit idle for 20 minutes to allow all
entrained air in oil to rise.
13. Repeat
Step 6
ten more times to ensure air removal
and inspect for leaks.
14. Disconnect, one at a time, hydraulic lines to floor
locks and purge each line. Remove as much oil as
possible from foot cylinder and refill manually
before reconnecting hydraulic line.
15. Level tabletop and raise table to half height.
Measure distance from tabletop to floor. Note this
dimension.
16. Allow sufficient time (minimum 30 minutes) to
ensure no downward drift. Measure distance
from tabletop to floor and compare to distance
recorded in
Step 15
.
17. Reassemble table and return to service.
7.36 LINEAR RAIL BEARING
REPLACEMENT
NOTE: Full tabletop weight is present on these cartridge
bearings. The bearing assembly attaches the sliding tabletop
seat section side member to the stationary tilt frame. A small
gap exists between the sliding section and the tilt frame. This
gap may be filled with a shim during bearing assembly
removal to ease the removal and replacement process.
1. Remove seat section and kidney elevator tabletops
, and
.
2. Place tabletop in "0" slide position.
3. Remove seat section side rail.
4. Remove cover plate (one additional screw) and
plastic cable cover.
5. Manually slide bearing assembly support to one
end of table seat section (until it reaches physical
stop) to expose eight bearing rail mounting screws
(four on upper rail and four on lower rail).
6. Remove eight visible bearing rail screws.
Hose
And Wire
Protection
Cap
Back
Sensor
Wiper
CAUTION – POSSIBLE EQUIPMENT
DAMAGE: Use care when separating the
bearing assembly from each side of the table
as the sliding tabletop side member is now
free-floating when the bearing assembly is
removed. Avoid bumping or misaligning the
slide potentiometer pulley
.
Summary of Contents for Cmax
Page 1: ......
Page 4: ......
Page 65: ......
Page 142: ...Page 3 of 14 ...
Page 148: ...Page 9 of 14 ...
Page 149: ...Page 10 of 14 ...
Page 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Page 164: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......