6
OPTIONS
STERIS 5085 SRT Surgical Tables are configured to anticipate
the future implementation of ACT (Advanced Control
Technology) software protocol. This interface would enable the
communication link to various Operating Room Control
Systems (ORCS) [not provided by STERIS]. Integration ORCS
software protocols would need to be validated by the integrator.
The ACT protocol would be validated at the time of
implementation.
PREVENTIVE MAINTENANCE
Customers are encouraged to contact STERIS concerning our
comprehensive preventive maintenance program. Under the
terms of this program, preventive maintenance, adjustments,
and replacement of worn parts are provided on a scheduled
basis to help ensure optimal equipment performance and help
avoid untimely or costly interruptions. STERIS maintains a
global staff of well equipped, factory-trained technicians to
provide these services, as well as expert repair services. Please
contact STERIS for details.
NOTES
1. Approximate Table Operating Weight: 850 lbs (386 kg).
2. Patient Weight Capacity:
•
1200 lb (544 kg)
load; with the ability to rotate and raise/
lower table but no tabletop articulation or slide.
•
1000 lb (454 kg)
load; with the ability to rotate and all
patient posturing but without tabletop slide.
•
600 lb (272 kg)
load; with the ability to rotate, all posturing,
transport feature and tabletop slide.
3. STERIS 5085 SRT Surgical Table accessories have specific
weight claims/capacities. Accessories may limit table patient
weight capacity.
4. A patient grounding post/potential equalization terminal
(male connector, DIN 42801) is provided with the table. The
female connector for patient grounding is not furnished by
STERIS.
5. WARNING – EXPLOSION HAZARD:
Table must not be used
in the presence of flammable anesthetics.
6. For further product information refer to Operator Manual
(413724-247) or contact STERIS corporation.
UTILITY REQUIREMENTS
Power Source:
Battery/Electric-Powered Table
Line Power Input:
100/120/220/230-240 V ac (Jumper Selectable), 1 Ø, 50/60 Hz,
4.0 Amp
Environmental Use Conditions:
Temperature: 50 - 104°F (10 - 40°C)
Relative Humidity (RH): 20 - 80%
CUSTOMER IS RESPONSIBLE FOR COMPLIANCE WITH
APPLICABLE LOCAL AND NATIONAL CODES AND
REGULATIONS.
The base language of this document is ENGLISH. Any translations
must be made from the base language document.