5
Patient Transport
can be actuated at any time provided the
floor locks are unlocked and the patient is in the correct
specified transport orientation (Transport mode). Patient
security is assured through the use of removable/lockable
patient safety barriers
and patient straps. Patient transport is
possible through a battery powered drive wheel. The user
controls the table steering. Deceleration and stopping is
accomplished by motor braking while moving forward. Forward
and reverse directional controls are incorporated into the
special
Drive Head Section
and allow for proportional speed
control (see picture). The patient transport feature also enables
the user to pivot a patient (around the center of the drive wheel)
for table repositioning.
The Patient Transport offers the following functions when
operating on a level surface:
• Table Speed:
Forward
≤
1.9 mph (
≤
3 km/h);
Reverse
≤
1.2 mph (
≤
2 km/h)
• Table Weight Capacity: 600 lb (272 kg)
• Stopping Distance (at 2 mph [ 3 kmh] with 600 lb patient):
≤
2’ (0.6 m) with motor braking in forward motion
Drive Head Section
is a cordless removable manually
operated table section containing the user control features for
the Patient Transport feature. The Drive Head Section (see
picture) incorporates two buttons: a drive engage/disengage
(when user pushes button the drive wheel engages) and
direction (forward and reverse). The Drive Head Section also
incorporates a center latch permitting the user to articulate the
head section from +60° to -90° in 10° increments. For added
security, the Drive Head Section has a locking feature
preventing section from inadvertent removal.
TABLE OPERATION
Press the desired position touch pad on the hand pendant or
optional foot control to position/articulate the STERIS 5085 SRT
Surgical Table tabletop (refer to Operator Manual).
The following pre-operative actions must be completed before
the table can be used:
•
Power activation:
Use battery power or plug the table
power cord into both table base and facility power
receptacle. Toggle Main Power Switch to ON for facility ac
power.
•
Turn the table on:
Press any button on the Hand Control
to turn table ON.
•
Lock the table:
Press the FLOOR LOCK button on the
Hand Control and then the LOCK button to order the floor
locks into the "locked" position.
Auto Limit Sensor™ control device (using the Intelligent Limit
Sensors) provides instant user feedback on conflicting tabletop
articulations when the table is positioned in standard
configuration and how to resolve the conflict.
TABLE MOTION
Refer to illustration Page 7 for the following ranges of table
motion:
• Flex/Reflex: 140°/100°±1°
• Slide Motion: HEAD: 9" (227 mm)
FOOT: 9" (227 mm) from center
• Height Range: 22.8 to 42.5" (581 to 1080 mm)
• Trendelenburg Range: 30°±1°
• Reverse Trendelenburg Range: 30°±1°
• Tilt Range: 20°±1°
• Back Section Motion: UP: 80°±1°
DOWN: 40°±1°
• Leg Section Motion: UP: 0°
DOWN: 105°±1°
• Kidney Elevator Mechanism: 4" (110 mm)
• Head Section (Manual): UP: 60°±1°
DOWN: 90°±1°, 10° increments
TREND
REVERSE
TREND
HEIGHT UP
HEIGHT DN
BACK UP
BACK DN
LEG UP
LEG DN
FLOOR
LOCK
STOP
CAUTION
Back-up mode
Auto Limit Sensors NOT available
USE WITH FOOT PUMP
SLIDE HEAD
SLIDE FOOT
KIDNEY UP
KIDNEY DN
FLOOR
UNLOCK
TILT LEFT
TILT RIGHT
Backup Hand Control
Drive Head Section
Drive Engage/Disengage Button
Direction Button
Hand Hold
Hand Hold
Stainless-Steel Side Rail