The pasteuriser operating criteria have been set for ambient room temperature of
25
˚C and incoming cold water from mains at 20 ˚C. Please ensure these criteria are
observed carefully as significantly higher temperatures will interfere with the
optimum running of the pasteuriser.
✓
Always ensure that the content of the probe bottle matches the temperature of
the bottles to be pasteurised to get accurate results.
✓
Only use waterproof labels suitable for prolonged exposure to hot & cold water.
✓
If labels comes off this can damage the pasteuriser.
✓
Please dry pasteuriser chamber after last cycle of the day.
Issue 5 4/12/2019
6
Do’s & Dont’s