THIS SECTION IS FOR ENGINEERING USE ONLY
❑
Suction Probe A Fault
Replace the probe
❑
Low Suction 1 A Fault
The refrigeration system is getting too cold. Follow the instructions in the Water Refrigeration
Section of this manual and check the bath temperature. If necessary adjust the cooling set point in
the setup section. Check refrigeration circuit has not lost refrigerant. A refrigeration engineer would
need to check this.
❑
Temp Probe Fault
One of the temperature probes is faulty. Replace the probe.
❑
Water Level High in Chill
Water level has risen in the cooling cycle. Check the neck level has not moved. If it has moved, the
inlet solenoid valve has failed. If the cold water supply pressure is very high it could be overfilling
before the solenoid valve is able to shut off in time. Partially close the cold water inlet tap supplying
the unit. Try to run a cycle, might need adjustment.
❑
Water Loss Fault
The water level in the bath has dropped during the cycle. Check the unit for leaks. Check that the
drain hose is installed correctly. Check the zone valve is working.
Issue 5 4/12/2019
32
Engineer Fault Finding
Summary of Contents for S180
Page 1: ...Operating Manual for S180 Pasteuriser Serial No This Manuals belongs to ...
Page 4: ...Please Add Specification Issue 5 4 12 2019 4 ...
Page 11: ...250ml 130ml 100ml 50ml Issue 5 4 12 2019 11 BOTTLE CHOICES Induction Sealer ...
Page 12: ...BOTTLE CHOICES 250ml 130ml 100ml 50ml Issue 5 4 12 2019 12 Pasteuriser Bottles ...
Page 15: ...HOME SCREEN Issue 5 4 12 2019 15 Pasteuriser Operation ...
Page 17: ...Bottle Probe Position Issue 5 4 12 2019 17 Pasteuriser Operation ...
Page 29: ...Data Logger Operation LIST VIEW GRAPH VIEW Issue 5 4 12 2019 29 Data Logger Operation ...
Page 40: ... Please add breakdown form Breakdown report form Issue 5 4 12 2019 40 ...
Page 41: ... Please add breakdown form Breakdown report form Issue 5 4 12 2019 41 ...
Page 42: ...Please add sequences of events chart Contact Us ...
Page 43: ...Contact Us Please add Graph ...