THIS SECTION IS FOR ENGINEERING USE ONLY
❑
Heater Overload
The Heater Overload F2 has tripped. Check the circuit to establish whether the heater has failed or
the re-set button has popped out, if there is a wiring fault or the overload is faulty.
❑
Chiller Overload
The Compressor overload F1 has tripped. Check the circuit to establish whether the compressor has
failed or the overload is faulty. Check hot & cold pipes are fitted the correct way around. Check
incoming cold water is below 25˚c
❑
Pump Overload
The Pump overload F4 or F5 has tripped. Check the circuit to establish whether the pump has failed, if
there is a loose connection or the overload is faulty.
❑
Low Gas Pressure
The refrigeration circuit may have lost refrigerant. A refrigeration engineer will be required to
diagnose this fault using the instructions in the Water Refrigeration System Section of the
maintenance manual. Check flow in tank during refrigeration cycle. If there is no flow check there is
no blockage in the water system. If there is no blockage the circulation pump has failed.
Engineer Fault Finding
Issue 5 4/12/2019
30
Summary of Contents for S180
Page 1: ...Operating Manual for S180 Pasteuriser Serial No This Manuals belongs to ...
Page 4: ...Please Add Specification Issue 5 4 12 2019 4 ...
Page 11: ...250ml 130ml 100ml 50ml Issue 5 4 12 2019 11 BOTTLE CHOICES Induction Sealer ...
Page 12: ...BOTTLE CHOICES 250ml 130ml 100ml 50ml Issue 5 4 12 2019 12 Pasteuriser Bottles ...
Page 15: ...HOME SCREEN Issue 5 4 12 2019 15 Pasteuriser Operation ...
Page 17: ...Bottle Probe Position Issue 5 4 12 2019 17 Pasteuriser Operation ...
Page 29: ...Data Logger Operation LIST VIEW GRAPH VIEW Issue 5 4 12 2019 29 Data Logger Operation ...
Page 40: ... Please add breakdown form Breakdown report form Issue 5 4 12 2019 40 ...
Page 41: ... Please add breakdown form Breakdown report form Issue 5 4 12 2019 41 ...
Page 42: ...Please add sequences of events chart Contact Us ...
Page 43: ...Contact Us Please add Graph ...