15. Insert the shaft and Magnet Lowering Device in the
gearbox hollow-shaft.
16. Screw the Magnet Lowering Device into the gearbox
frame.
17. Put the drive unit upright (use care and appropriate
lifting devices where required) and raise the shaft by
operating the Magnet Lowering Device.
STRONG MAGNETIC FIELDS HAZARD:
The
Sterimixer impeller and magnetic rotor include
strong magnets with the associated risk of pinching
fingers and permanently damaging magnetic cards
(e.g. credit cards) if being close to these compo-
nents.
HAZARD FOR WEARERS OF CARDIAC PACE-
MAKER:
The Sterimixer impeller and magnetic
rotor include strong magnets thus personnel
equipped with pacemaker shall not handle these
components.
Fix the magnet rotor on the shaft with the magnet
rotor screw.
18. For final tightening of the magnet rotor and Magnet
Lowering Device mounting screw, put the drive unit
on its side (use care and appropriate lifting devices
where required) and tighten as shown in the picture
below.
8. FITTING THE OPTIONAL SENSOR KIT TO
THE MAGNET LOWERING DEVICE
1. Remove the M3x3 screws from the Magnet Lowering
Device housing.
www.steridose.com
Installation & Operation
|
7