METHOD OF USE
In order to begin using your new accessory, follow these
instructions:
1. Insert the connectors in the corresponding inputs on your
MUL-285. Make sure to respect the polarity (+ Red, - Black).
2. Select the type of output depending on the type of
measurement you wish to make.
3. Adjust the multimeter to the VAC function and select a range.
4. Pull the trigger in order to open the measurement hooks and
place them on the cavle you wish to measure
5. Make sure that the hooks are
fi
rmly closed around the
conductor and check the reading on the multimeter’s LCD
screen.
6. When you are done needing this accessory, power the
multimeter off and remove the connectors.
SPECIFICATIONS
The precision is speci
fi
ed for a period of a year after calibration
and at a temperature between 18°C and 28°C with a relative
humidity of 80%.
Current Range:
0.1A – 400A AC rms
Voltage output level:
10 mV/A: 1mV AC for 0.1 Amps
AC
1 mV/A: 1 mV AC for 1 Amp AC
Operation voltage:
600 V CAT II for IES 1010-1
Operation temperature:
0°C to 50°C < 80% relative
humidity
Storage temperature:
-20°C to 60°C
Sensor type:
Induction sensor for AC
current
Maximum output impedance:
45 ohms
Largest hook opening:
29 mm
Maximum conductor size:
28 mm
Dimensions:
64 x 143 x 30 mm
Weight:
185g