Fuse replacement
-To prevent electric shock, shaken by
arcing or damage to person or the meter,
use only the speci
fi
ed fuse according to
the following procedure.
1. Turn the meter off and remove all
connections from the same.
2. Remove the two rubber bands and the
screws from the deck; separate meter
cover.
3. Remove the fuse arm and install a
new one with the same speci
fi
cations
(F10AL250B) of the sample; make sure
the fuse is
fi
rmly attached to the arm.