29
9 Technical
Data
Inhalation Anesthesia Unit MPG-Class 2b, Model Registration Number: 11 / M-01491
Typ: B Protection Class: 1
U:
230 V AC
P:
80 VA (with ventilator)
Dimensions (with Ventilator ABVA)
Width:
500 mm
Height: 1350
mm
Depth:
630 mm
Dimensions (without Ventilator ABVA)
Width:
500 mm
Height: 1240
mm
Depth:
630 mm
Connections (threaded) for Gas Types:
O
2
:
M 12 x 1
AIR:
M 20 x 1.5
N
2
O:
M 14 x 1
Fresh Gas:
M 16 x 1.5
Circuit pressure
5 bar +/- 0.5 bar
Measuring range for measuring hoses:
O
2
(fine)
:
0 to 2 litres per minute
(roughly)
:
2 to 15 litres per minute
AIR
:
0 to 15 litres per minute
N
2
O
:
0 to 15 litres per minute
Precision of Measuring tubes:
+/- 10% of the respective end scale value. For integrated micro measuring range +/- 10%
of the measuring endscale value (under standard conditions 20ºC and 1.013 bar)
O
2
deficiency signal:
Trigger
pressure:
2.8
bar
Duration of acoustic warning signal
7 seconds
N
2
O Block:
Trigger
pressure:
2.0
bar