background image

5

TROUBLESHOOTING: CAUSES AND SOLUTIONS

PROBLEM 

POTENTIAL CAUSE 

POTENTIAL SOLUTION

No power to Desk

 

No power at wall outlet 

Check outlet with another device;

 

 

check circuit breakers/fuses/wiring.

 

Power cable not plugged 

Confi rm proper cable engagement into wall outlet.

 

into wall outlet

 

Power cable not plugged 

Check that Power Cable is fully seated into Control Box.

 

into Control Box 

(see Fig. 1 on page 6)

 

Faulty Power Cable 

Inspect the Power Cable for damage; replace any damaged cables.

 

Desk wiring not properly 

Check that all cables are connected per the Assembly Directions.

 

assembled 

When fi rst plugged in, the Control Box will click twice to confi rm

  

proper 

connection.

Digital Controller does

 

No power at wall outlet 

See above.

not work or illuminate 

Control Box has wrong 

Control Box fi rmware must be 1.9 or later for Digital/Preset Controller

 fi 

rmware 

compatibility. 

(see Fig. 4 on page 6)

Desk will go neither

 

No Power 

See above.

 

Desk is severely 

Maximum load capacity is 103 kg. Overloading the Desk

 

overloaded 

could lead to damage that would not be covered by warranty.

 

 

ERROR CODE E24/25/26, E48/49 = Over Current

 

 

Tip: When calculating lifting capacity, subtract the weight

 

 

of the worksurface

 

Control Box requires 

Perform initialization procedure. 

(see Common Procedures)

 

reset 

Lower Desk all the way down, and perform synchronization

  

procedure. 

(see Common Procedures)

 

Damaged pins inside 

Ensure the Desk is in ‘Standby’ mode. (>20 seconds of inactivity)

 

multi-pin connectors 

Unplug each multi-pin connection point and inspect the pins in the

 

(Low-Voltage Cable / 

Control Box connector.

 

connector) 

Are they all straight, and making good contact?

 

 

If not, try using a push-pin to straighten them. 

(see Fig. 2 on page 6)

 

 

(Otherwise, replace necessary components with undamaged versions.)

 

Damaged Lifting Column 

Carefully check the condition of the Low-Voltage Cable of each

 

Low-Voltage Cable 

Lifting Column. 

(see Fig. 3 on page 6)

 

 

If any Low-Voltage Cables are damaged, the entire Lifting Column

 

 

must be replaced.

 

Faulty Lifting Column 

To prevent further damage, the Control Box will the whole Desk

 

 

down if one Lifting Column is faulty. Test Lifting Columns 

(see step 4

 

 

of Troubleshooting Steps)

, and replace any faulty Lifting Columns.

 

Faulty Controller (or) 

Swap with Controller known to work (and/or) check operation.

 faulty 

controller

 

receptacle in Control Box

 

Outdated Digital 

Control Box fi rmware before version 1.9 is not compatible with the

 

Controller software 

Digital/Preset Controller. 

(see Fig. 4 on page 6)

 

 

Replace with a new Control Box with fi rmware 1.9 or later.

 

Faulty Control Box 

Swap with Control Box known to work.

 

Damaged pins inside 

Unplug Power Cable from wall. Unplug each multi-pin connection

 

multi-pin connectors 

point, and inspect the pins. Are they all straight, and making good

 

(Low-Voltage Cable / 

contact? If not, try using a push-pin to straighten them.

 connector) 

(See Fig. 2 on page 6)

 (Otherwise, replace necessary components

 

 

with undamaged versions.)

Digital Controller does
not work or illuminate

Desk will go neither
up nor down

Desk will go neither
up nor down

Summary of Contents for Ology Bench

Page 1: ...ox controls all aspects of Desk motion and synchronizes motion of the individual Lifting Columns The Control Box will shut down the entire Desk if a fault is detected For example if one Lifting Column is binding or trapped it will draw significantly more current than the other Lifting Column and the Control Box will shut down the entire Desk to prevent further damage The Controller is the user inte...

Page 2: ...damage to them Step 4 Test Lifting Columns for proper function and replace any faulty Lifting Columns To do this create a bench testing setup using known good components for example from your troubleshooting parts kit Use these known good components to test the Lifting Columns of the suspect Desk as follows A Verify the function of the bench test setup by connecting all the known good components i...

Page 3: ...ty C Using the original Power Cable and Controller swap the known good Control Box with the original from the non functioning Desk If it stops working the Control Box is faulty connector Be sure to go through all of the steps above to fully identify all faulty components There could be more than one COMMON PROCEDURES POWER SAVING CONTROL BOX The Control Box automatically powers down into Standby m...

Page 4: ... positions 1 2 and the Up button simultaneously for 10 seconds 2 Display shows S7 3 Press and hold the Up button until the display reads S0 4 Press S the unit is reset to factory settings SYNCHRONIZING INITIALIZING LIFTING COLUMNS Operate the Desk all the way down to the lowest position and release the Down button Press the Down button again and hold After a few seconds the Desk will visibly move ...

Page 5: ...seconds of inactivity multi pin connectors Unplug each multi pin connection point and inspect the pins in the Low Voltage Cable Control Box connector connector Are they all straight and making good contact If not try using a push pin to straighten them see Fig 2 on page 6 Otherwise replace necessary components with undamaged versions Damaged Lifting Column Carefully check the condition of the Low ...

Page 6: ...lumn Test Lifting Columns per step 4 of the Troubleshooting Steps is not smooth and replace any faulty Lifting Columns Be sure to test all Lifting Lifting Columns jerk Columns Why Because another Lifting Column may have the same or hop while raised problem at a different height lowered Lifting Columns are not First ensure that glides are properly adjusted and the Desk is level parallel with one an...

Page 7: ...d at power off HW initialization FW driver initialization Wait until the HOT error message is no longer displayed May need to wait for entire remaining 18 minutes of 20 minute duty cycle period Try soft reset unplug wait 5 sec plug back in if this does not solve the issue re initialize the Control Box Verify function of Control Box via troubleshooting steps Replace Control Box if faulty Control bo...

Page 8: ...er com Steelcase youtube com SteelcaseTV 7 2017 2017 Steelcase Inc All rights reserved All specifications subject to change without notice Trademarks used herein are the property of Steelcase Inc or of their respective owners Part No 1076390001 Rev A ...

Reviews: