background image

2

READ THE ASSEMBLY DIRECTIONS AND USER GUIDE

Many times, problems can be the result of improper assembly. Reference the Assembly Directions
document to ensure the Desk is assembled correctly. If so, reference the troubleshooting steps throughout
this document.

TROUBLESHOOTING PARTS KIT

Having a spare part that is known to work will quickly lead to accurate identifi cation of faulty components.
Without this, it can be little more than guesswork.
Steelcase strongly recommends that every Dealer maintain a small kit of parts for troubleshooting purposes.

TROUBLESHOOTING STEPS

Step 1:

 

Check for power at the wall outlet; ensure the Desk is properly assembled, and all cables are

 

securely connected. Upon powering up the Desk, the Control Box will click twice to indicate

 

normal operation and verify that connections are properly made. If it clicks only once, or if it

 

clicks 3 or more times, there is a fault in the system.

Step 2:

   Check all connections of Lifting Column Low-Voltage Cables. Check the connection points on

 

the Control Box: are any of the pins in the connectors damaged or not making contact?

 

(see Fig. 2 on page 6)

Step 3:

 

Obtain a Control Box, Digital / Preset Controller, and Power Cable that are known to work.

Caution! If the Lifting Column is assembled to a Desk, fi rst fl ip the Desk upside-down. This will allow
full and free motion of the Lifting Columns, and will prevent any further damage to them.

Step 4:

 

Test Lifting Columns for proper function, and replace any faulty Lifting Columns. To do this,

 

create a bench-testing setup using known good components, for example from your

 

troubleshooting parts kit. Use these known good components to test the Lifting Columns of

 

the suspect Desk as follows:

 

A.  Verify the function of the bench-test setup by connecting all the known-good components,

 

 

including Lifting Columns, synchronize the Lifting Columns (see Common Procedures), and

 

 

check for proper function.

 

B.  Disconnect power to the Control Box, and connect the Lifting Columns of the suspect Desk to

 

 

the bench-test setup. If the Lifting Columns of the suspect Desk now operate normally using the

 

 

bench-test setup, then the Lifting Columns of the suspect Desk are good. Move on to Step 5.

 

C.  If the Lifting Columns of the suspect Desk do not operate normally, observe the error code

 

 

shown in the display of the Digital / Preset Controller of the bench-test setup. The following

 

 

error codes may indicate a fault with the Lifting Column, its Low-Voltage Cable, or the cable

  

connector:

 

 

1.  E12 / E13 / E14: Defect on channel 1 / 2 / 3.

 

 

2.  E24 / E25 / E26: Overcurrent on motor M1 / M2 / M3

 

 

3.  E36 / E37 / E38: Plug detection in motor socket M1 / M2 / M3

 

D.  Using error code E12 as an example…error code E12 may indicate a fault with the Lifting

 

 

Column plugged into channel 1 of the Control Box. To verify this, fi rst disconnect power to the

 

 

Control Box, and then swap the Lifting Column low-voltage connectors between channels

 

 

1 and 2, so that Lifting Column 1 is now plugged into 2,

 

 

and 2 is plugged into 1. Re-connect the power.

Summary of Contents for Ology Bench

Page 1: ...ox controls all aspects of Desk motion and synchronizes motion of the individual Lifting Columns The Control Box will shut down the entire Desk if a fault is detected For example if one Lifting Column is binding or trapped it will draw significantly more current than the other Lifting Column and the Control Box will shut down the entire Desk to prevent further damage The Controller is the user inte...

Page 2: ...damage to them Step 4 Test Lifting Columns for proper function and replace any faulty Lifting Columns To do this create a bench testing setup using known good components for example from your troubleshooting parts kit Use these known good components to test the Lifting Columns of the suspect Desk as follows A Verify the function of the bench test setup by connecting all the known good components i...

Page 3: ...ty C Using the original Power Cable and Controller swap the known good Control Box with the original from the non functioning Desk If it stops working the Control Box is faulty connector Be sure to go through all of the steps above to fully identify all faulty components There could be more than one COMMON PROCEDURES POWER SAVING CONTROL BOX The Control Box automatically powers down into Standby m...

Page 4: ... positions 1 2 and the Up button simultaneously for 10 seconds 2 Display shows S7 3 Press and hold the Up button until the display reads S0 4 Press S the unit is reset to factory settings SYNCHRONIZING INITIALIZING LIFTING COLUMNS Operate the Desk all the way down to the lowest position and release the Down button Press the Down button again and hold After a few seconds the Desk will visibly move ...

Page 5: ...seconds of inactivity multi pin connectors Unplug each multi pin connection point and inspect the pins in the Low Voltage Cable Control Box connector connector Are they all straight and making good contact If not try using a push pin to straighten them see Fig 2 on page 6 Otherwise replace necessary components with undamaged versions Damaged Lifting Column Carefully check the condition of the Low ...

Page 6: ...lumn Test Lifting Columns per step 4 of the Troubleshooting Steps is not smooth and replace any faulty Lifting Columns Be sure to test all Lifting Lifting Columns jerk Columns Why Because another Lifting Column may have the same or hop while raised problem at a different height lowered Lifting Columns are not First ensure that glides are properly adjusted and the Desk is level parallel with one an...

Page 7: ...d at power off HW initialization FW driver initialization Wait until the HOT error message is no longer displayed May need to wait for entire remaining 18 minutes of 20 minute duty cycle period Try soft reset unplug wait 5 sec plug back in if this does not solve the issue re initialize the Control Box Verify function of Control Box via troubleshooting steps Replace Control Box if faulty Control bo...

Page 8: ...er com Steelcase youtube com SteelcaseTV 7 2017 2017 Steelcase Inc All rights reserved All specifications subject to change without notice Trademarks used herein are the property of Steelcase Inc or of their respective owners Part No 1076390001 Rev A ...

Reviews: