background image

3

 

E.  If the error code now reads E13, indicating a fault on channel 2, then the fault moved with

 

 

the Lifting Column. Therefore, the problem must reside with the Lifting Column 1, which was

 

 

moved from channel 1 to 2.

 

F. 

If the error code instead continues to read E12, then both Lifting Columns may be faulty.

 

G.  For further verifi cation, repeat this process by swapping out the suspect Desk’s Lifting

 

 

Columns with the Lifting Columns from your troubleshooting parts kit, to zero in on which

 

 

Lifting Column is faulty, if not both. (If the suspect Desk’s Lifting Columns function normally

 

 

in the bench-test setup, move on to Step 5.)

 

Any faulty Lifting Columns must be replaced.

 

There are no fi eld-serviceable parts inside the Lifting Column.

Between each of the following steps, be sure to disconnect the Control Box from the power source
for at least 5 seconds to power down the internal processor.

Step 5:

 

If there are no error codes to indicate faults with the Lifting Columns or their Low-Voltage Cables

 

or connectors, progressively swap out parts in the bench-test setup as follows:

 

A.  Swap the known-good Power Cable with the original from the non-functioning Desk.

 

 

If it stops working, the Power Cable is faulty.

 

B.  Using the original Power Cable, swap the known good Controller with the original from

 

 

the non-functioning Desk. If it stops working, the Controller is faulty.

 

C.  Using the original Power Cable and Controller, swap the known good Control Box with

 

 

the original from the non-functioning Desk. If it stops working, the Control Box is faulty.

  

connector:

Be sure to go through all of the steps above to fully identify all faulty components.
There could be more than one!

COMMON PROCEDURES

POWER-SAVING CONTROL BOX:

 

·

The Control Box automatically powers down into ‘Standby’ mode after 20 seconds of inactivity.

 

·

During ‘Standby’ mode, the Control Box only consumes 0.3W watt of power.

 

·

With the Control Box in ‘Standby’ mode, the Desk can lose power without affecting the synchronization

 

of the Lifting Columns. However, a power loss still allows the Control Box to power down and ‘Soft Reset.’

 (see 

below)

DUTY CYCLE:

 

·

In order to prevent excessive heat buildup and avoid damage to the system, the Control Box limits the

 

Up-Down function of the Desk to a maximum of 1 minute of continuous operation in a 10-minute timeframe

 

(10% duty cycle).

 

·

If the 10% duty cycle is exceeded, the Control Box may still allow limited movement, for example 3 seconds

 

of movement after 10 seconds of rest. If further movement is desired, allow the Desk to cool off before

 

making further adjustments.

 

·

If the Desk does not function at all after long periods of operation, allow the Desk to cool off, until the

 

Control Box has been idle for the entire duty cycle period.

 

·

If the Control Box overheats, the display of the Digital / Preset Controller will read ‘HOT.’ Simply wait until

 

the Control Box cools off, and ‘HOT’ is no longer displayed, before attempting further adjustments.

Summary of Contents for Ology Bench

Page 1: ...ox controls all aspects of Desk motion and synchronizes motion of the individual Lifting Columns The Control Box will shut down the entire Desk if a fault is detected For example if one Lifting Column is binding or trapped it will draw significantly more current than the other Lifting Column and the Control Box will shut down the entire Desk to prevent further damage The Controller is the user inte...

Page 2: ...damage to them Step 4 Test Lifting Columns for proper function and replace any faulty Lifting Columns To do this create a bench testing setup using known good components for example from your troubleshooting parts kit Use these known good components to test the Lifting Columns of the suspect Desk as follows A Verify the function of the bench test setup by connecting all the known good components i...

Page 3: ...ty C Using the original Power Cable and Controller swap the known good Control Box with the original from the non functioning Desk If it stops working the Control Box is faulty connector Be sure to go through all of the steps above to fully identify all faulty components There could be more than one COMMON PROCEDURES POWER SAVING CONTROL BOX The Control Box automatically powers down into Standby m...

Page 4: ... positions 1 2 and the Up button simultaneously for 10 seconds 2 Display shows S7 3 Press and hold the Up button until the display reads S0 4 Press S the unit is reset to factory settings SYNCHRONIZING INITIALIZING LIFTING COLUMNS Operate the Desk all the way down to the lowest position and release the Down button Press the Down button again and hold After a few seconds the Desk will visibly move ...

Page 5: ...seconds of inactivity multi pin connectors Unplug each multi pin connection point and inspect the pins in the Low Voltage Cable Control Box connector connector Are they all straight and making good contact If not try using a push pin to straighten them see Fig 2 on page 6 Otherwise replace necessary components with undamaged versions Damaged Lifting Column Carefully check the condition of the Low ...

Page 6: ...lumn Test Lifting Columns per step 4 of the Troubleshooting Steps is not smooth and replace any faulty Lifting Columns Be sure to test all Lifting Lifting Columns jerk Columns Why Because another Lifting Column may have the same or hop while raised problem at a different height lowered Lifting Columns are not First ensure that glides are properly adjusted and the Desk is level parallel with one an...

Page 7: ...d at power off HW initialization FW driver initialization Wait until the HOT error message is no longer displayed May need to wait for entire remaining 18 minutes of 20 minute duty cycle period Try soft reset unplug wait 5 sec plug back in if this does not solve the issue re initialize the Control Box Verify function of Control Box via troubleshooting steps Replace Control Box if faulty Control bo...

Page 8: ...er com Steelcase youtube com SteelcaseTV 7 2017 2017 Steelcase Inc All rights reserved All specifications subject to change without notice Trademarks used herein are the property of Steelcase Inc or of their respective owners Part No 1076390001 Rev A ...

Reviews: