Removing and Installing the Calibration Chamber
To separate the calibration chamber, you require ZU 0754 or ZU 0740 Service Set (see Accessories).
We recommend the ZU 0746 and ZU 0747 mounting aids for proper mounting of the gaskets and
scraper rings.
1) Remove the screws
(screwdriver TX25).
How to handle the mounting aids
is described in the respective user
manual.
4) To re-install the split calibration
chamber, screw the parts together
using plier and face pin spanner
wrench and secure them with
screws.
Note:
The calibration chamber parts
must be firmly screwed together
(until the stop is reached) before
it can be secured with the two
screws.
2) Position a plier and loosen the
thread of the split calibration
chamber using a face pin spanner
wrench.
3) Completely screw off the split
calibration chamber. Now, the
gaskets are accessible and can be
checked and replaced if required.
Use the ZU 0746 and ZU 0747
mounting aids for mounting the
gaskets and scraper rings.