STATUS SCIENTIFIC CONTROLS
Installation, Commissioning & Routine Gas Testing
FGD10B Pellistor Gas Detector
TD18/008
Issue:
10
Change Note:
1904
Page 20
4.5.
Calibration / Configuration Keypad
The FGD10B sensor can be calibrated via a purpose designed keypad. The
keypad allows the user to carry out the following:
1)
Calibrate the sensor.
2)
Calibrate the 4 to 20 mA loop.
3)
View the current gas level.
4)
View the sensor raw data for diagnostic purposes.
Connect the keypad into the FGD10B as shown below:
The display will give the following messages:
SSCL
Company
- - - -
88:88
Segment test
SSCL
Manufacturer
1.0.3
Firmware version
- - - -
0.0
Gas reading
Note: if the display is continually showing - - - - then the instrument is not
communicating with the keypad.