4.2.3.2 curnt1(3) CURRENT OUTPUT
(RETRANSMISSION ) BOARD
The current retransmission board provides a range of current output options. If fitted,
the following menu will be available from the Output menu.
4.2.3.2.1 SPAN ( Output current span )
Span is the current range at which the output board is to operate.
The options are:
4-20mA
( Output current will vary from 4 - 20mA )
0-20mA
( Output current will vary from 0 - 20mA )
0-10mA
( Output current will vary from 0 - 10mA )
OUTPUT MENU