If the Voltage Output Board ( Bridge Excitation ) is to be fitted in the unit by the user,
it is necessary to use this entry to calibrate the option.
Connect up the Voltage output board to the
Sensor input board as shown below.
Press
ENTER,
to
commence automatic
calibration.
Note if the display reads either
'siglo'
or
'sighi'
recalibrate the output card again.
On completion, ESCAPE back to the Process Variable display mode. This stores the
calibration information into non-volatile memory.
4.2.5.5 rESEt ( Reset all parameters to default )
Pressing ENTER on this option resets all programmable parameters to their default
values.
CALIBRATION MENU
PROGRAMMING
Page 67
Summary of Contents for DM4000U
Page 3: ...GETTING STARTED Page 3 ...
Page 7: ...INSTALLATION Page 7 ...
Page 25: ...USER GUIDE Page 25 ...