36 DiagCORE Respiratory Panel 2 Instructions for Use
Further data about the assay is available, depending on the operator’s access rights, through the
tabs at the bottom of the screen (e.g., amplification plots and test details).
A report with the assay data can be exported to an external USB storage device. Insert the USB
storage device into one of the USB ports of the DiagCORE
®
Analyzer and press Save Report in the
bottom bar of the screen. This report can be exported later at any time by selecting the test from
the View Result List.
The report can also be sent to the printer by pressing Print Report in the bottom bar of the screen.
8.1.1
Viewing amplification curves
To view test amplification curves of pathogens detected, press the
Amplification Curves tab.
Amplification Curves screen (Pathogens).
Details about the tested pathogens and controls are shown on the left, and the amplification
curves are shown in the center.
NOTE
If User Access Control is enabled on the DiagCORE
®
Analyzer the Amplification
Curves screen is only available for operators who have access rights.
Press the Pathogens tab on the left side to display the plots corresponding to the tested pathogens.
Press on the pathogen name to select which pathogens are shown in the amplification plot. It is
possible to select single, multiple, or no pathogens. Each pathogen in the selected list will be
assigned a color corresponding to the amplification curve associated with this pathogen.
Unselected pathogens will be shown in gray.
The corresponding C
T
and endpoint fluorescence values are shown below each pathogen name.
Press the Controls tab on the left side to view the controls and select which controls are shown in
the amplification plot. Press the circle next to the control name to select or deselect it.