26 DiagCORE Respiratory Panel 2 Instructions for Use
3.
Manually write the sample information or place a sample information label on the top of the
DiagCORE
®
cartridge. Make sure that the label is properly positioned and does not block the
lid opening.
4.
Open the sample lid of the main port on the front of the cartridge.
5.
Open the tube with the sample to be tested. Use the supplied transfer pipette to draw up fluid
to the third fill line on the pipette (300 µL).