32 DiagCORE Respiratory Panel 2 Instructions for Use
NOTE
The lid of the cartridge entrance port will close automatically after 30
seconds if no cartridge is positioned in the port. If this occurs, repeat the
procedure starting with step 5.
9.
While the test is running, the time remaining in the run is displayed on the screen.
10.
After the test run is completed, the Eject screen will appear and the Module Status Bar will show
the test result as one of the following options:
TEST COMPLETED
The test was completed successfully.
TEST FAILED
An error occurred during the test.
TEST CANCELED
The user canceled the test.
NOTE
If the test fails, refer to the DiagCORE Analyzer User Manual Troubleshooting
section for possible reasons and instructions on how to proceed.
11.
Press
Eject to remove the cartridge and dispose of it as biohazardous waste in accordance
with all national, state, and local health and safety regulations and laws.
The cartridge should be removed when the cartridge entrance port opens and ejects the
cartridge. If the cartridge is not removed after 30 seconds, it will be automatically moved back
into the Analyzer, and the lid of the cartridge entrance port will close. If this occurs, press Eject
to open the lid of the cartridge entrance port again and then remove the cartridge.
NOTE
Used cartridges must be discarded.
It is not possible to re-use cartridges for tests for which the execution was
started but then subsequently canceled by the operator, or for which an
error was detected.
12.
After the cartridge has been ejected, the Results Summary screen will appear. Please refer to
section Interpretation of Results for further details. To run another test, press Run Test.