PS300 Series High Voltage Power Supplies
Commands
3 – 7
*SAV
i
Save Con
fi
guration
The
*SAV
command stores the present setup as se ing
i
.
i
may
range from 1 to 9.
Example:
*SAV 3
ILIM(?) {
x
}
Limit
Current
The ILIM command sets (queries) the value of the current limit
(to
x
), in amperes. ILIM may be set from 0 to 105 % of full scale.
Example:
ILIM 120E-6; ILIM?
1.20E-4
ITRP(?) {
x
}
Trip Current
The
ITRP
command sets (queries) the value of the current trip
(to
x
), in amperes. ITRP may be set from 0 to 105 % of full scale.
Example:
ITRP? 5.25E-4
SMOD(?) {
i
}
Rear Programming Mode
The
SMOD? command sets (queries)
the VSet se ing mode. SMOD 0
means that the voltage value is controlled by the front-panel
se ing, while SMOD 1 indicates that the output is controlled by
the rear-panel VSET voltage control input.
Note that changing the SMOD value while the high voltage is ON
causes the high voltage to be switched OFF.
Example:
SMOD 1
TCLR
Clear Trips
The
TCLR command
clears any voltage or current trips.
Example:
TCLR
Summary of Contents for PS355
Page 4: ...Stanford Research Systems PS300 Series High Voltage Power Supplies Contents ii ...
Page 16: ...Stanford Research Systems PS300 Series High Voltage Power Supplies Instrument Overview 1 6 ...
Page 22: ...Stanford Research Systems PS300 Series High Voltage Power Supplies Guide to Operation 2 6 ...
Page 38: ...Stanford Research Systems PS300 Series High Voltage Power Supplies Remote Operation 3 16 ...
Page 42: ...Stanford Research Systems PS300 Series High Voltage Power Supplies Troubleshooting 4 4 ...