5 Assembly 

a)  

Working environment requirement

 

1)

 

Room temperature: 0-40

Maximum comparatively humidity: 85%.

 

2

 

High temperature will result in mildew, dew and even ruinous instrument.

 

3)  

Avoid from the dust room. When it is not in use, please cover the 

microscope with dust cover.

 

4

 

Please place the microscope in a stable situation without any 

surging. 

b)  

Check the Input voltage: be sure the power supply voltage is accordant with the 

nominal input voltage which is signed outside the microscope, or it will bring a serious 

damage to the microscope.

 

c)

 

 

Lamp

 

1.

 

 The lamp has been well adjusted and checked before the microscope 

leaves factory.

   

2

 

.

 The lamp housing parts become extremely hot during operation. Take care 

not to burn yourself.

 

3.

 

 Lamp replacement

The lamp has its standard service life. When it has 

expired, a lamp replacement is necessary.set the main switch to off state, 

disconnect the power cord from the wall outlet, then allow the old bulb to 

cool before replacing the bulb with a new of the designated type .Open the 

lamp holder on the bottom of microscope to do the replacement, the surface 

of the new lamp bulb should be clean and free of fingerprint or dirt, which will 

decrease the brightness or even explode the bulb

 . 

Note: Must not missing the grounded screw

 

 

   

       

Donot touch the lamp`s glass part with your hands,when sitting shoud 

take gloves or cover the lamp with protector. Fingerprint or dirt on the 

surface of the lamp should be cleaned with the clean cloth with dipping 

alcohol,which will decrease the brightness or even explode the bulb.

 

d)  

Eyepiece

 

Please insert the eyepieces into the eyepiece tubes.(There is small screw in 

the eyepiece ,which can make the eyepiece can not be pull out arbitrarily 

when be screwed down.

 

e)

 Objective

 

Lower the stage to its lower limit, then screw down the objectives into the 

nosepiece (from low to high)

f)

 Condenser

 

he condenser has been adjusted well before delivery

g)

 Collector

 

The collector has been adjusted well before delivery. Frame of the 

collector`s pedestal is screw thread can taken by contrarotate

   

Note

It cannot be taken down when the outfit is LED. 

Summary of Contents for STM-2010 Series

Page 1: ...ries To ensure the safety obtain optimum performance and to familiarize yourself fully with the use of this microscope it is recommended strongly that you study this manual thoroughly before operating...

Page 2: ...2 STM 2010D Eyepiece Objective Monocular Head Coarse Focusing Knob Fine Focusing Knob Brightness Adjustment Knob Main Switch Stage Condenser Collector...

Page 3: ...mination general biologic pathologic and bacteriological studies clinical investigations and classroom demonstrations Designed in a very modern way what can take convenient and safe to your operations...

Page 4: ...X 2 5 Conjugate Distance 195mm 2 6 Stage STM 2010 D E Double Layers Mechanical Stage 116mmX125mm STM 2010 C Plain Stag with Attachable Mechanical ruler 110 120mm 60 30 mm STM 2010 A B Plain Stage with...

Page 5: ...h Iris Diaphragm Rack and Pinion Adjustment Nosepiece Triple Nosepiece Quadruple Nosepiece Plain Stage with Standard Standard Standard Optional Optional Optional Optional Standard Standard Standard St...

Page 6: ...rved arm use C shaped has sprightly lines Also with upper focus stop the coarse focusing adjustment adopts dovetail guide which make focusing adjustment smooth and comfortable c Base The base is desig...

Page 7: ...efore replacing the bulb with a new of the designated type Open the lamp holder on the bottom of microscope to do the replacement the surface of the new lamp bulb should be clean and free of fingerpri...

Page 8: ...into coincidence Generally speaking the rate of objective increase the location of the condenser need to move up Aperture is design for adjust NA and not for adjust brightness Usually aperture is open...

Page 9: ...nd stain in the field of view There are stains on the lens include condenser objective eyepiece and collector Clean it up There are stains on the specimen Clean it up The position of the condenser is...

Page 10: ...nosepiece is not in the right position Turn it to the clicked position The image is a little yellow Not use the blue color filter Use the blue filter The brightness is not enough The size of the apert...

Page 11: ...ower cord The bulb is not inserted Insert it correctly The bulb burns out Replace it The lamp burns out suddenly Use an unstandard lamp The voltage is too high Use the specified lamp to replace if the...

Reviews: