The Zeiss LSM880 is a laser scanning confocal microscope with full confocal and multiphoton capability.  The 
confocal detectors are spectrally resolved and can be used with either the standard (single  photon) excitation 
lines or the multiphoton excitation lines.  The non-descanned (NDD) detectors can only be used with multiphoton 
excitation. 

 

I

NSTRUMENT OVERVIEW 

 

AND USER RESOURCES

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F

ULL 

T

URN ON 

S

EQUENCE 

(F

IRST ONE ON

1)  If the computer is off, turn it on and let it boot. 

2)  Turn on the main switch. 

3)  Turn on the Systems/PC switch.  (When 

touchscreen has booted, adjust heating and 
CO2 under Microscope->Inkubation if 
applicable.) 

4)  Turn on the Components switch.  Wait for the 

RTC to boot (see under table). 

5)  Log in to the computer. 

6)  Start Zen and select 

Start System

 Button.   

7)  Turn on laser lines you will be using.  

(If you are 

using multiphoton excitation, turn on both lines 
on even if you are just using one.  Note that the 
Ar and multiphoton lines take a while to turn on.  
You can follow the warm-up progress by 
selecting the laser and the Laser Properties 
below)

 

 

F

ULL 

T

URN OFF 

S

EQUENCE 

(L

AST ONE ON

9)  Save your data and transfer it to the fileshare. 

8)  Remove your sample and clean off your 

objective.  

(If using incubation, turn temp and 

CO2 regulation off and close CO2 tank using 
main gray valve only.) 

7)  Turn off lasers. 

6)  Exit Zen

5)  Log out of the computer. 

4)  Turn off the Components switch. 

3)  Turn off the Systems/PC switch. 

2)  Wait a few minutes for the Ar laser to cool.  

Turn off the main switch. 

1)  Do not turn off computer. 

 

 

 

 

 

 

P

ARTIAL 

T

URN OFF 

S

EQUENCE

 

(Someone else is scheduled after your time slot) 
 

9)  Save your data and transfer it to the fileshare. 

8)  Remove your sample and clean off your 

objective.  

(If using incubation, turn temp and 

CO2 regulation off and close CO2 tank using 
gray handle only.) 

7)  Turn off lasers. 

6)  Exit Zen.

 

5)  Log out of the computer.  

 

 

 

P

ARTIAL 

T

URN ON 

S

EQUENCE

 

(Microscope already on) 

5)  Log in to the computer. 

6)  Start Zen and select 

Start System

 Button. 

7)  Turn on laser lines you will be using.  

(If you are 

using multiphoton excitation, turn on both lines 
on even if you are just using one.  Note that the 
Ar and multiphoton lines take a while to turn on.  
You can follow the warm-up progress by 
selecting the laser and the Laser Properties 
below)

 