12.1 Close the water tap and remove the electric
plug from the socket, then disconnect the raw
water hose from the cartridge.
CAUTION:
When opening, water may leak out of the
cartridge. Make sure, that this does not impair any
other systems in the vicinity.
12.2 Remove the cartridge with conductivity meter
from the bracket.
12.3 Unscrew the conductivity meter from the ex-
hausted cartridge.
12.4 Fit the conductivity meter on the fresh cartridge
12
Change of cartridge
stakpure mixed-bed ion exchanger
DS 450
DS 450
Max. flow rate
50 l/h
50 l/h
Capacity at 10°GSG (total dissolved solids, 10° dH GSG=12.5° e)
425 l
425 l
Pure water quality
0.1 – 20 μS/cm
0.1 – 20 μS/cm
Max. water temperature
30° C
30° C
Electric connection
230 V / 50-60 Hz, 10 VA
115 V / 50-60 Hz, 10 VA
Height x diameter, incl. conductivity meter
100 x 600 mm
100 x 600 mm
Depth
115 mm
115 mm
Weight
3 kg
3 kg
Article no.
11042500
11042502
Replacement cartridge (2 pcs.)
11042700
11042700
Height x diameter
100 x 435 mm
100 x 435 mm
13
Technical Data DS 450
Designation:
Mix-bed ion exchanger
Type:
Stakpure DS 450
Article No.:
11042500 / 11042502
14
EEC – �eclaration o� con�ormit�
Applicable EEC Directives:
EEC- Directive low voltage
EEC- Directive electromagnetic compatibility
EMV 2004/108/EG
Applied standard:
55011 und 61000
Niederahr, June 2014
_____________________________________________________
Manufacturer/CE-Representative: Leo Trumm
and fit the combination in the bracket. Connect
the raw water hose to the fresh cartridge. Only
hand tighten the hose nipple.
12.5 Plug in the power plug. The system is now
ready for use again. Please note: When no
water flows through the cartridge, reionization
effects could cause the conductivity meter to
show incorrect values. Please do not use the
water produced until the flow of it through the
measuring cell brings the conductivity meter
display to a correct value.