background image

8

Measurement Operation

DC Voltage Measurement (see Figure 2)

Warning

To avoid injury to you or damage to the multimeter from electric 
shock, please do not attempt to measure voltages higher than 500V 
although readings may be obtained.

The DC voltage ranges are: 200 mV, 2000 mV, 20V, 200V and 500V. 
To measure DC voltage, connect the multimeter as follows:

1.  Insert the red test lead into the VΩmA terminal and the black test 

lead into the COM terminal.

2.  Set the multimeter knob to the appropriate measurement position 

in V     range.

3.  Connect the test leads to the object being measured.
4.  The measured value shows on the display.
5.  1.5V, 9V, 12V is for battery testing. The meter puts a small load on 

the battery to get a more accurate reading.

Note:

1.  If the value of voltage to be measured is unknown, use the 

maximum measurement position (500V) and reduce the range by 
step until a satisfactory reading is obtained.

2.  The LCD displays “1” indicating the existing selected range is 

overload. It is required to select a higher range in order to obtain  
a correct reading.

3. 

In each range, the meter has input impedance of 10 MΩ. This effect 

can cause measurement errors in high impedance circuits. If the circuit  

impedance is less than or equal to 10kΩ, the error is negligible 

(0.1% or less).

4.  When DC voltage measurement has been completed, disconnect 

the connection between the testing leads and the circuit under test.

AC Voltage Measurement (see figure 2)

Warning

To avoid injury to you or damage to the meter from electric shock, please do not attempt to measure voltages higher than 
500V RMS although readings may be obtained.

The AC voltage measurement positions are 200V and 500V. To measure AC voltage, connect the multimeter as follows:

1.  Insert the red test lead into the VΩmA terminal and the black test lead into the COM terminal.

2.  Set the knob of multimeter to an appropriate measurement position in V~ range.
3.  Connect the test leads to the object being measured.
4.  The measured value shows on the display, which is effective value of sine wave (mean value response).

Note

1.  If the value of voltage to be measured is unknown, use the maximum measurement position (500V) and reduce the range 

the range by step until a satisfactory reading is obtained.

2.  The LCD displays “1” indicating the existing selected range is overload is overload; it is required to select a higher range 

in order to obtain a correct reading.

3.  In each range, the meter has an input impedance of 10 MΩ. This loading effect can cause measurement errors in high 

impedance circuits. If the circuit impedance is less than or equal to 10kΩ, the error is negligible (0.1% or less).

4.  When AC voltage measurement has been completed, disconnect the connection between the testing leads and the circuit 

under test.

Figure 2

Summary of Contents for DSTS

Page 1: ...Model DSTS User Manual Variable Temperature Digital Solder and Testing Station...

Page 2: ...ll reach as high as 400 C 752 F do not touch metallic parts near the soldering tip when the power switch is on 7 A fire may result if the appliance is not used with care therefore do not leave the app...

Page 3: ...d probe for most testing applications except high amperage 7 Negative socket for black probe Power Supply 8 Coarse adjustment voltage control 9 Negative wire jack 10 Ground jack if needed 11 Positive...

Page 4: ...y elements will be switched on briefly The electronic system will automatically turn on to the set temperature and display the current value Display and temperature setting 1 Display for actual temper...

Page 5: ...er or electrical specifications before using the meter 3 Do not apply more than the rated voltage as marked on the multimeter between the terminals or between any terminal and grounding 4 The rotary s...

Page 6: ...play 2 Display backlight button 3 HOLD button 4 Function selector knob 5 Positive 10A socket for red probe for high amperage testing only 6 Positive socket for red probe for most testing applications...

Page 7: ...000m 10 Battery Type One 9V Battery NEDA 1604 6F22 or 006P 11 Battery Deficiency Display 12 Negative reading Display 13 Overload Display 1 14 Accuracy a reading b digits see charts below guarantee for...

Page 8: ...voltage measurement has been completed disconnect the connection between the testing leads and the circuit under test AC Voltage Measurement see figure 2 Warning To avoid injury to you or damage to th...

Page 9: ...ined 2 When current measurement has been completed disconnect the connection between the testing leads and the circuit under test Measuring Resistance see Figure 4 Warning To avoid damage to the multi...

Page 10: ...t test General Service 1 Periodically wipe the case with a damp cloth and mild detergent Do not use abrasive materials or solvents 2 Clean the terminals with a cotton swab with detergent as dirt or mo...

Page 11: ...and or installation 2 This warranty does not cover cosmetic damage due to misuse or neglect This includes paint damage scratches cracks or other superficial marks related to improper use 3 Failures ar...

Page 12: ...Stahl Tools Last Revised 9 15 2017 stahltools com...

Reviews: