SSCOR S-SCORT new-Sentinel 2315 Operating Instructions And Maintenance Manual Download Page 5

General Specifications

CHARACTERISTICS                                     SPECIFICATIONS

Size

17”L x 9”H x 6“W
(43.2cm L x 22.9cm H x 15.2cm W)

Weight

8.5 lbs. (4.1 Kgs.)

Vacuum Pump

12V DC oil-less diaphragm. 3.0A
Exceeds -525mmHg
Clinical Airflow >= 30LPM

Variable Regulator (7)

Controls vacuum from < -50 to >= -525mmHg

Gauge (8)

Calibrated in mmHg. Color-coded.

Power Source:  Battery (DC Power)

Rechargeable Sealed Lead Acid.
45 (±10%) minute running time.

Switch (5)

On/Off rocker switch

Collection Canister (13)

1200cc/ml  SSCOR part #48041

Patient tubing

Vinyl tubing 9/32”ID, 72”L

SSCOR part #43200

Suction Tip

HI-D

® 

“Big Stick

®”

 SSCOR part #44241

Trouble Shooting

PROBLEM                             POSSIBLE CAUSE               CORRECTIVE ACTION

Does not function when

Battery discharged

Reconnect to battery

switch is on

charger

Power source connected

Fuse blown

Refer to Bio-Med

indicator light not lit

Bad connection

Refer to Bio-Med

No suction when pump is

Vacuum line (10) loose

Tighten

running

Canister(13) damaged

Replace

Canister lid (13) loose

Tighten

Catheter thumb vent (6) is
open

Occlude

Debris has collected in

Refer to maintenance

pump

section of this manual

– 5 –

Summary of Contents for S-SCORT new-Sentinel 2315

Page 1: ...RUCTIONS AND MAINTENANCE MANUAL S SCORT new Sentinel Model 2315 SSCOR INC 11064 Randall Street Sun Valley CA 91352 USA Telephone 1 818 504 4054 800 434 5211 Fax 1 818 504 6032 www sscor com Email marketing sscor com ...

Page 2: ...pump must be reconnected to the battery charger after each use and remain connecteduntilneeded 6 Do not use in the presence of flammable agents or anesthetics 7 Theshut offvalveinthecanisterlidwillclosedownwhenthecanisterisfilledwithfluids 8 Installanewcanisterbeforetestingforvacuumover 300mmHgtominimizethepossibility of implosion which can occur when a canister is aged or damaged 9 SSCOR s batter...

Page 3: ...y reduce the available capacity High temperatures may cause deformation of the battery case and damage the battery Sealed lead acid batteries can easily be maintained to permit proper operation of the equipment To protect the battery after each procedure turn the unit off put the unit on charge andalwaysstorethepumpatroomtemperature Theonlywaytoassurethebatteryhasfunctional capacity evenifitisindi...

Page 4: ...llection canister 13 6 Press the ON OFF switch 5 on control panel to start vacuum pump 7 Occlude the patient tube turn the vacuum regulator 7 clockwise to the stop See Caution Notice page 2 number 8 8 Observe the gauge 8 The vacuum reading should rise to 300mmHg from the zero setting in 3 seconds It should max out at approximately 525mmHg Un occlude the patient tubing 9 Allow the unit to run for 1...

Page 5: ...ectionCanister 13 1200cc ml SSCOR part 48041 Patienttubing Vinyltubing9 32 ID 72 L SSCOR part 43200 SuctionTip HI D Big Stick SSCOR part 44241 Trouble Shooting PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION Does not function when Battery discharged Reconnect to battery switch is on charger Power source connected Fuse blown Refer to Bio Med indicator light not lit Bad connection Refer to Bio Med No sucti...

Page 6: ...assumes nor authorizes any representative or otherpersonstoassumeforit anyotherliabilityinconnectionwiththesaleofSSCORproducts Thiswarrantygivesyouspecificlegalrightsandyoumayalsohaveotherrightsthatvaryfrom jurisdictiontojurisdiction Forcountrieswhereminimumwarrantytermsaredeterminedby statute the warranty term is the longer of the statutory period or the term listed above Batteries disposableitem...

Page 7: ... the unlikely event that fluids may have reached the exhaust filter remove the two screws holding the filter door in place see page 8 remove the soiled filter and replace with a new filter SSCOR P N 90160 Do not reuse any single use disposable parts do not submerge the device into any liquid this will void the warranty and cause the device to malfunction Decontamination Use personal protective equ...

Page 8: ...2V Sealed Lead Acid REMOVE MAIN CHASSIS L Vaccum pump Return to SSCOR for SCREWS service Do not attempt to service the pump A 1 screw under the canister holder M Exhaust barb B 2 screws in the back of the handle N Vaccum barb B 3 screws in the back of the main chassis C It should not be necessary to open the exhaust door unless fluids have entered the system Slowlypulltheunitapartatthehandletoobse...

Reviews: