General Description

The S-SCORT® new-Sentinel is a portable, 12V DC battery operated suction pump for
constant suctioning during resuscitation emergencies.  It is powered by sealed lead acid
batteries, capable of driving the unit for 45 minutes 

±

 10%.  The batteries are recharged by a

115V AC to 13.8V DC Dual Mode Charger (included with the unit).  A sealed lead acid
battery is a very stable and reliable battery.  Many factors can affect the life of a battery:

Leaving a unit switched on after there is no longer enough power to run the pump
can cause a battery to deep discharge.  This can reduce the life of, or destroy the battery.

Failing to charge a battery for an extended period of time will also cause the battery to go
into deep discharge.

Low temperatures may reduce the available capacity.

High temperatures may cause deformation of the battery case and damage the battery.

Sealed lead acid batteries can easily be maintained to permit proper operation of the
equipment.   To protect the battery, after each procedure turn the unit off, put the unit on charge
and always store the pump at room temperature.  The only way to assure the battery has functional
capacity, even if it is indicated the battery is fully charged, is to perform the weekly check suggested
on page 4 of this operations manual.

The new-Sentinel is designed to provide instant, effective suctioning, independent of external
sources of power and can be pre-set to be activated immediately upon reaching the distressed
patient.  Suction power is controllable for those instances when full power would be considered
harmful to the patient.  The S-SCORT new-Sentinel is equipped with a disposable collection
canister which features a bacterial filter to screen airborne particulates and a mechanical shut-off
valve to prevent fluid overflow.

Description of Symbols

– 3 –

Symbol

Meaning

Location

Type B Medical Equipment              Serialized Label

Serialized Label                               Separate Collection for
                                                        Electronic Equipment

Serialized Label                               Caution - Consult
                                                        Accompaning Documents

Summary of Contents for S-SCORT new-Sentinel 2315

Page 1: ...RUCTIONS AND MAINTENANCE MANUAL S SCORT new Sentinel Model 2315 SSCOR INC 11064 Randall Street Sun Valley CA 91352 USA Telephone 1 818 504 4054 800 434 5211 Fax 1 818 504 6032 www sscor com Email marketing sscor com ...

Page 2: ...pump must be reconnected to the battery charger after each use and remain connecteduntilneeded 6 Do not use in the presence of flammable agents or anesthetics 7 Theshut offvalveinthecanisterlidwillclosedownwhenthecanisterisfilledwithfluids 8 Installanewcanisterbeforetestingforvacuumover 300mmHgtominimizethepossibility of implosion which can occur when a canister is aged or damaged 9 SSCOR s batter...

Page 3: ...y reduce the available capacity High temperatures may cause deformation of the battery case and damage the battery Sealed lead acid batteries can easily be maintained to permit proper operation of the equipment To protect the battery after each procedure turn the unit off put the unit on charge andalwaysstorethepumpatroomtemperature Theonlywaytoassurethebatteryhasfunctional capacity evenifitisindi...

Page 4: ...llection canister 13 6 Press the ON OFF switch 5 on control panel to start vacuum pump 7 Occlude the patient tube turn the vacuum regulator 7 clockwise to the stop See Caution Notice page 2 number 8 8 Observe the gauge 8 The vacuum reading should rise to 300mmHg from the zero setting in 3 seconds It should max out at approximately 525mmHg Un occlude the patient tubing 9 Allow the unit to run for 1...

Page 5: ...ectionCanister 13 1200cc ml SSCOR part 48041 Patienttubing Vinyltubing9 32 ID 72 L SSCOR part 43200 SuctionTip HI D Big Stick SSCOR part 44241 Trouble Shooting PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION Does not function when Battery discharged Reconnect to battery switch is on charger Power source connected Fuse blown Refer to Bio Med indicator light not lit Bad connection Refer to Bio Med No sucti...

Page 6: ...assumes nor authorizes any representative or otherpersonstoassumeforit anyotherliabilityinconnectionwiththesaleofSSCORproducts Thiswarrantygivesyouspecificlegalrightsandyoumayalsohaveotherrightsthatvaryfrom jurisdictiontojurisdiction Forcountrieswhereminimumwarrantytermsaredeterminedby statute the warranty term is the longer of the statutory period or the term listed above Batteries disposableitem...

Page 7: ... the unlikely event that fluids may have reached the exhaust filter remove the two screws holding the filter door in place see page 8 remove the soiled filter and replace with a new filter SSCOR P N 90160 Do not reuse any single use disposable parts do not submerge the device into any liquid this will void the warranty and cause the device to malfunction Decontamination Use personal protective equ...

Page 8: ...2V Sealed Lead Acid REMOVE MAIN CHASSIS L Vaccum pump Return to SSCOR for SCREWS service Do not attempt to service the pump A 1 screw under the canister holder M Exhaust barb B 2 screws in the back of the handle N Vaccum barb B 3 screws in the back of the main chassis C It should not be necessary to open the exhaust door unless fluids have entered the system Slowlypulltheunitapartatthehandletoobse...

Reviews: