SSCOR S-SCORT new-Sentinel 2315 Operating Instructions And Maintenance Manual Download Page 4

Getting Acquainted

Charging the Battery

1. Connect the dual mode battery charger (1) to a grounded wall receptacle.
2. Connect the female plug of the battery charger (1A) to the facing side of the unit (1A).
3. Check the control panel (2).  The "power source connected" red LED (4) indicates a good

connection between the new-Sentinel and the charger.

4. Check the LED’s on the battery charger (1).  "Power On" (21) indicates a good connection

to the AC power source.  "Fast Charge" (22) will light when the battery is charging and go
out when the battery is fully charged.  When the "Fast Charge" (22) light is out, you can
test the charging mode by flicking the switch (5) to the "On" position and back "Off".  The
"Fast Charge" (22) light should come on immediately.

TEST FOR VACUUM.  RUN THE FOLLOWING TEST WEEKLY TO ENSURE
PROPER PERFORMANCE OF THE DEVICE.

5. Attach patient tubing to patient port (12) on collection canister (13).
6. Press the ON/OFF switch (5) on control panel to start vacuum pump.
7. Occlude the patient tube, turn the vacuum regulator (7) clockwise to the stop.  See Caution

- Notice, page 2 number 8.

8. Observe the gauge (8). The vacuum reading should rise to -300mmHg from the zero setting

in 3 seconds.  It should max out at approximately >= -525mmHg.  Un-occlude the patient
tubing.

9.  Allow the unit to run for 15 minutes on DC power.  If the unit stops or slows during the 15

minutes, it is possible the battery capacity has been depleted.  It is time to replace the
battery.

10.  If the unit is still running at full power after 15 minutes, adjust the regulator to the desired

setting, turn the device off and put it back on charge.

NOTE A:  If the vacuum does not reach -525mmHg, check for a leak in the system i.e. vacuum

tubing (10), barb connections (11,12) or loose canister lid (13).  Refer internal vacuum
problems to qualified personnel.

NOTE B: Reconnect the charger to the unit as soon as possible after each use by attaching the

charging plug (1A) to the receptacle on the unit (1A).

– 4 –

Summary of Contents for S-SCORT new-Sentinel 2315

Page 1: ...RUCTIONS AND MAINTENANCE MANUAL S SCORT new Sentinel Model 2315 SSCOR INC 11064 Randall Street Sun Valley CA 91352 USA Telephone 1 818 504 4054 800 434 5211 Fax 1 818 504 6032 www sscor com Email marketing sscor com ...

Page 2: ...pump must be reconnected to the battery charger after each use and remain connecteduntilneeded 6 Do not use in the presence of flammable agents or anesthetics 7 Theshut offvalveinthecanisterlidwillclosedownwhenthecanisterisfilledwithfluids 8 Installanewcanisterbeforetestingforvacuumover 300mmHgtominimizethepossibility of implosion which can occur when a canister is aged or damaged 9 SSCOR s batter...

Page 3: ...y reduce the available capacity High temperatures may cause deformation of the battery case and damage the battery Sealed lead acid batteries can easily be maintained to permit proper operation of the equipment To protect the battery after each procedure turn the unit off put the unit on charge andalwaysstorethepumpatroomtemperature Theonlywaytoassurethebatteryhasfunctional capacity evenifitisindi...

Page 4: ...llection canister 13 6 Press the ON OFF switch 5 on control panel to start vacuum pump 7 Occlude the patient tube turn the vacuum regulator 7 clockwise to the stop See Caution Notice page 2 number 8 8 Observe the gauge 8 The vacuum reading should rise to 300mmHg from the zero setting in 3 seconds It should max out at approximately 525mmHg Un occlude the patient tubing 9 Allow the unit to run for 1...

Page 5: ...ectionCanister 13 1200cc ml SSCOR part 48041 Patienttubing Vinyltubing9 32 ID 72 L SSCOR part 43200 SuctionTip HI D Big Stick SSCOR part 44241 Trouble Shooting PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION Does not function when Battery discharged Reconnect to battery switch is on charger Power source connected Fuse blown Refer to Bio Med indicator light not lit Bad connection Refer to Bio Med No sucti...

Page 6: ...assumes nor authorizes any representative or otherpersonstoassumeforit anyotherliabilityinconnectionwiththesaleofSSCORproducts Thiswarrantygivesyouspecificlegalrightsandyoumayalsohaveotherrightsthatvaryfrom jurisdictiontojurisdiction Forcountrieswhereminimumwarrantytermsaredeterminedby statute the warranty term is the longer of the statutory period or the term listed above Batteries disposableitem...

Page 7: ... the unlikely event that fluids may have reached the exhaust filter remove the two screws holding the filter door in place see page 8 remove the soiled filter and replace with a new filter SSCOR P N 90160 Do not reuse any single use disposable parts do not submerge the device into any liquid this will void the warranty and cause the device to malfunction Decontamination Use personal protective equ...

Page 8: ...2V Sealed Lead Acid REMOVE MAIN CHASSIS L Vaccum pump Return to SSCOR for SCREWS service Do not attempt to service the pump A 1 screw under the canister holder M Exhaust barb B 2 screws in the back of the handle N Vaccum barb B 3 screws in the back of the main chassis C It should not be necessary to open the exhaust door unless fluids have entered the system Slowlypulltheunitapartatthehandletoobse...

Reviews: