General Description

The SSCOR Sentinel is a portable, 12V DC battery operated suction pump for constant
suctioning during resuscitation emergencies.  It is powered by sealed lead acid batteries,
capable of driving the unit for 45 minutes ± 10%.  The batteries are recharged by a 115V AC to
13.8V DC Dual Mode Charger (included with the unit).  A sealed lead acid battery is a very
stable and reliable battery.  Many factors can affect the life of a battery:

Leaving a unit switched on after there is no longer enough power to run the pump
can cause a battery to deep discharge.  This can reduce the life of, or destroy the
battery.

Failing to charge a battery for an extended period of time will also cause the battery to
go into deep discharge.

Low temperatures may reduce the available capacity.

High temperatures may cause deformation of the battery case and damage the battery.

Sealed lead acid batteries can easily be maintained to permit proper operation of the
equipment. To protect the battery, after each procedure turn the unit off, put the unit on
charge and always store the pump at room temperature.  The only way to assure the
battery has functional capacity, even if it is indicated the battery is fully charged, is to perform
the battery test suggested on page 4 of this operations manual.

The Sentinel is designed to provide instant, effective suctioning, independent of external
sources of power and can be pre-set to be activated immediately upon reaching the distressed
patient.  Suction power is controllable for those instances when full power would be considered
harmful to the patient.  The SSCOR Sentinel is equipped with a disposable collection canister
which features a bacterial filter to screen airborne particulates and a mechanical shut-off valve
to prevent fluid overflow.

Description of Symbols

– 3 –

Symbol

Meaning

Location

Type B Medical Equipment             Serialized Label

Serialized Label                              Separate Collection for
                                                        Electronic Equipment

Serialized Label                              Caution - Consult
                                                        Accompaning Documents

Summary of Contents for 2315

Page 1: ...NS AND MAINTENANCE MANUAL SSCOR Sentinel Model 2315 SSCOR INC 11064 Randall Street Sun Valley CA 91352 USA Telephone 1 818 504 4054 800 434 5211 Fax 1 818 504 6032 www sscor com Email info sscor com techsupport sscor com ...

Page 2: ...ith these operating instructions before this device is used 5 The suction pump must be reconnected to the battery charger after each use and remain connected until needed 6 Do not use in the presence of flammable agents or anesthetics 7 The shut off valve in the canister lid will close down when the canister is filled with fluids 8 Install a new canister before testing for vacuum over 300mmHg to m...

Page 3: ...e capacity High temperatures may cause deformation of the battery case and damage the battery Sealed lead acid batteries can easily be maintained to permit proper operation of the equipment To protect the battery after each procedure turn the unit off put the unit on charge and always store the pump at room temperature The only way to assure the battery has functional capacity even if it is indica...

Page 4: ...e proper performance of the device Confirm the power cord is supplying power to the device Check the power indicator light on the control panel Remove the power cord and run the unit from its internal DC battery Check the vacuum by occluding the patient tube and set the vacuum regulator to the maximum vacuum setting Observe the regulator gauge The vacuum reading should rise to 300mmHg from zero in...

Page 5: ...ection Canister 13 1200cc ml SSCOR part 48041 Patient tubing Vinyl tubing 9 32 ID 72 L SSCOR part 43200 Suction Tip HI D Big Stick SSCOR part 44241 Trouble Shooting PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION Does not function when Battery discharged Reconnect to battery switch is on charger Power source connected Fuse blown Refer to Bio Med indicator light not lit Bad connection Refer to Bio Med No ...

Page 6: ... or implied and of all other obligations or liabilities on SSCOR s part and SSCOR neither assumes nor authorizes any representative or other persons to assume for it any other liability in connection with the sale of SSCOR products This warranty gives you specific legal rights and you may also have other rights that vary from jurisdiction to jurisdiction For countries where minimum warranty terms ...

Page 7: ...ble parts do not submerge the device into any liquid this will void the warranty and cause the device to malfunction Disinfection Use personal protective equipment such as gloves a smock and face and eye protection when handling units that are suspected to be contaminated Part Cleaning and Disinfecting Collection Canister Disposable item re use not permitted Use new canister for each patient Patie...

Page 8: ...lips head fastening screws are located in back of the unit REMOVE MAIN CHASSIS SCREWS A 1 screw under the canister holder B 2 screws in the back of the handle B 3 screws in the back of the main chassis C It should not be necessary to open the exhaust door unless fluids have entered the system G Ground is connected between the receptacle and the pump bracket H DC receptacle I Battery Bracket J Batt...

Reviews: