SSCOR 2315 Operating Instructions And Maintenance Manual Download Page 7

– 7 –

Maintenance

Sanitation

As soon as possible after use, the single use disposable canister, patient tubing and catheter
should be discarded according to local / regional / national requirements for the disposal of
hazardous waste materials.  Clean the exterior of the SSCOR suction unit using a mild deter-
gent and clear water by dampening a clean lint free cloth.  Rinse using clear water and another
damp clean lint free cloth to remove any detergent residue.

NOTE: The hydrophobic filter in the canister helps to ensure that no moisture or particulate

matter reaches the inside of the device.  When fluids fill the canister, the positive (me-
chanical float) shutoff valve closes immediately, shutting the vacuum port off so as to
prevent fluid from contacting the pump. The filter has been tested by the manufacturer
(Bemis) to screen out aerosolized microorganisms and particulate matter at a bacterial
efficiency rating of 99.99% DOP.  The canister also has sidewall gradation marks start-
ing at 100 ml/cc and at every 50 ml/cc up to 1200 ml/cc indicating the fill level of the
canister.

In the unlikely event that fluids may have reached the vacuum pump, read the disinfection
section.  Your engineering department will have to open the unit to check the condition of the
pump.
Do not reuse any single use disposable parts; do not submerge the device into any liquid, this
will void the warranty and cause the device to malfunction.

Disinfection

Use personal protective equipment such as gloves, a smock, and face and eye protection when
handling units that are suspected to be contaminated.

Part

Cleaning and Disinfecting

Collection Canister

Disposable item, re-use not permitted.  Use new canister
for each patient.

Patient Tubing

Disposable item, re-use not permitted.  Use new patient
tubing for each patient.

HI-D

®

 Stick

Disposable item, re-use not permitted.  Use new HI-D Stick
for each patient.

Vacuum Pump

Wipe with damp cloth or disinfectant wipe.  Sterilization not
permitted.  Vacuum pump should be replaced if
contaminated.

Chassis

Wipe with damp cloth or disinfectant wipe.  Sterilization not
permitted.

Caution:

 Disconnect the unit from any power source prior to cleaning the unit.

Disinfect the unit using a mild surface disinfectant, such as a 10:1 mixture of water and bleach.
The unit is designed to suction contaminated fluids, which should be removed from the system
immediately after use.   In the unlikely event that fluids may have reached the vacuum pump, your
engineering department will have to open the unit to check the condition of the pump.  When
cleaning the interior of the chassis, disconnect the battery from the wiring harness. The only
forseeable way fluids may reach the vacuum pump is the filter in the canister has been compro-
mised or bypassed.

For technical assistance, call (800) 434-5211 or +1 (818) 504-4054.

Summary of Contents for 2315

Page 1: ...NS AND MAINTENANCE MANUAL SSCOR Sentinel Model 2315 SSCOR INC 11064 Randall Street Sun Valley CA 91352 USA Telephone 1 818 504 4054 800 434 5211 Fax 1 818 504 6032 www sscor com Email info sscor com techsupport sscor com ...

Page 2: ...ith these operating instructions before this device is used 5 The suction pump must be reconnected to the battery charger after each use and remain connected until needed 6 Do not use in the presence of flammable agents or anesthetics 7 The shut off valve in the canister lid will close down when the canister is filled with fluids 8 Install a new canister before testing for vacuum over 300mmHg to m...

Page 3: ...e capacity High temperatures may cause deformation of the battery case and damage the battery Sealed lead acid batteries can easily be maintained to permit proper operation of the equipment To protect the battery after each procedure turn the unit off put the unit on charge and always store the pump at room temperature The only way to assure the battery has functional capacity even if it is indica...

Page 4: ...e proper performance of the device Confirm the power cord is supplying power to the device Check the power indicator light on the control panel Remove the power cord and run the unit from its internal DC battery Check the vacuum by occluding the patient tube and set the vacuum regulator to the maximum vacuum setting Observe the regulator gauge The vacuum reading should rise to 300mmHg from zero in...

Page 5: ...ection Canister 13 1200cc ml SSCOR part 48041 Patient tubing Vinyl tubing 9 32 ID 72 L SSCOR part 43200 Suction Tip HI D Big Stick SSCOR part 44241 Trouble Shooting PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION Does not function when Battery discharged Reconnect to battery switch is on charger Power source connected Fuse blown Refer to Bio Med indicator light not lit Bad connection Refer to Bio Med No ...

Page 6: ... or implied and of all other obligations or liabilities on SSCOR s part and SSCOR neither assumes nor authorizes any representative or other persons to assume for it any other liability in connection with the sale of SSCOR products This warranty gives you specific legal rights and you may also have other rights that vary from jurisdiction to jurisdiction For countries where minimum warranty terms ...

Page 7: ...ble parts do not submerge the device into any liquid this will void the warranty and cause the device to malfunction Disinfection Use personal protective equipment such as gloves a smock and face and eye protection when handling units that are suspected to be contaminated Part Cleaning and Disinfecting Collection Canister Disposable item re use not permitted Use new canister for each patient Patie...

Page 8: ...lips head fastening screws are located in back of the unit REMOVE MAIN CHASSIS SCREWS A 1 screw under the canister holder B 2 screws in the back of the handle B 3 screws in the back of the main chassis C It should not be necessary to open the exhaust door unless fluids have entered the system G Ground is connected between the receptacle and the pump bracket H DC receptacle I Battery Bracket J Batt...

Reviews: