Rev. N (03/14)

Caution-Notice

1. SSCOR suction units are not designed or intended for use in extended procedures

that require prolonged  high vacuum/low airflow applications, as is the case in wound
drainage or endoscopic use or in any other procedure that produces high vacuum
levels within an occluded system for an extended period of time.  Turn the suction
unit off when it is not in use.

2. Federal law restricts this device to sale, distribution, and use by, or on the order of a

physician, emergency medical technician, or other medical practitioner. For use by
medical personnel trained in suctioning techniques and in the use of medical suction
equipment.

3. This manual is restricted to the discussion of the use and maintenance of this device.

It does not attempt to discuss professional techniques in suctioning procedures.

4. Operator should be thoroughly familiar with these operating instructions before this

device is used.

5. The suction pump must be reconnected to the battery charger after each use and

remain connected until needed.

6. Do not use in the presence of flammable agents or anesthetics.

7. The shut-off valve in the canister lid will close down when the canister is filled with

fluids.

8. Install a new canister before testing for vacuum over -300mmHg to minimize the

possibility of implosion, which can occur when a canister is aged or damaged.

9. SSCOR's battery charger charges batteries only.  The battery charger will not run the

pump.

10.  This unit complies with electromagnetic compatibility standards.  Any person who

connects external equipment to signal input, signal output, or other connectors has
formed a system and is therefore responsible for the system to comply with the
requirements of IEC 60601-1-2.  If in doubt, contact a qualified technician or your local
representative.

SSCOR Sentinel

 

Model 2315 Series

Battery operated portable suction pump

©2014 HI-D and Big Stick are registered trademarks of SSCOR, Inc.

– 2 –

Summary of Contents for 2315

Page 1: ...NS AND MAINTENANCE MANUAL SSCOR Sentinel Model 2315 SSCOR INC 11064 Randall Street Sun Valley CA 91352 USA Telephone 1 818 504 4054 800 434 5211 Fax 1 818 504 6032 www sscor com Email info sscor com techsupport sscor com ...

Page 2: ...ith these operating instructions before this device is used 5 The suction pump must be reconnected to the battery charger after each use and remain connected until needed 6 Do not use in the presence of flammable agents or anesthetics 7 The shut off valve in the canister lid will close down when the canister is filled with fluids 8 Install a new canister before testing for vacuum over 300mmHg to m...

Page 3: ...e capacity High temperatures may cause deformation of the battery case and damage the battery Sealed lead acid batteries can easily be maintained to permit proper operation of the equipment To protect the battery after each procedure turn the unit off put the unit on charge and always store the pump at room temperature The only way to assure the battery has functional capacity even if it is indica...

Page 4: ...e proper performance of the device Confirm the power cord is supplying power to the device Check the power indicator light on the control panel Remove the power cord and run the unit from its internal DC battery Check the vacuum by occluding the patient tube and set the vacuum regulator to the maximum vacuum setting Observe the regulator gauge The vacuum reading should rise to 300mmHg from zero in...

Page 5: ...ection Canister 13 1200cc ml SSCOR part 48041 Patient tubing Vinyl tubing 9 32 ID 72 L SSCOR part 43200 Suction Tip HI D Big Stick SSCOR part 44241 Trouble Shooting PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION Does not function when Battery discharged Reconnect to battery switch is on charger Power source connected Fuse blown Refer to Bio Med indicator light not lit Bad connection Refer to Bio Med No ...

Page 6: ... or implied and of all other obligations or liabilities on SSCOR s part and SSCOR neither assumes nor authorizes any representative or other persons to assume for it any other liability in connection with the sale of SSCOR products This warranty gives you specific legal rights and you may also have other rights that vary from jurisdiction to jurisdiction For countries where minimum warranty terms ...

Page 7: ...ble parts do not submerge the device into any liquid this will void the warranty and cause the device to malfunction Disinfection Use personal protective equipment such as gloves a smock and face and eye protection when handling units that are suspected to be contaminated Part Cleaning and Disinfecting Collection Canister Disposable item re use not permitted Use new canister for each patient Patie...

Page 8: ...lips head fastening screws are located in back of the unit REMOVE MAIN CHASSIS SCREWS A 1 screw under the canister holder B 2 screws in the back of the handle B 3 screws in the back of the main chassis C It should not be necessary to open the exhaust door unless fluids have entered the system G Ground is connected between the receptacle and the pump bracket H DC receptacle I Battery Bracket J Batt...