Preven ve Care 

Observe the following maintenance rou ne to ensure readiness at any  me: 
1. When the SSCOR aspirator is not in use, keep ba eries on con nuous charge. 
2. Test the SSCOR aspirator at regular intervals; See page 6. 
3. Make sure the SSCOR aspirator is always clean and ready for use. 
4. If the procedure produced an excessive quan ty of 

fl

uids, check the vacuum line (7) for evidence of 

moisture. If the vacuum line between the pump and canister is moist, it is possible that 

fl

uids have reached 

the vacuum pump. See Disinfec on Instruc ons below. 
5. For technical assistance, call (800) 434

5211 or interna onal +1 818

504

4054. For replacement parts 

and accessories, please visit shop.sscor.com. 
Note : No part requires lubrica on and lubricants should not be used. 
 

Sanita on 

As soon as possible a er use, the single use disposable canister, pa ent tubing and catheter should be  
discarded according to local / regional / na onal requirements for the disposal of hazardous waste  
materials. Clean using a mild detergent and if necessary disinfect with a mild disinfectant such as 10 to 1 
bleach and rinse using clear water to remove any residue. 
If the exhaust 

fi

lter is soiled, remove the two screws holding the 

fi

lter door in place (see page 8), remove 

the soiled 

fi

lter and replace with new 

fi

lter (SSCOR part number 90160). 

Do not reuse any single use disposable parts; do not submerge the device into any liquid, this will void the 
warranty and cause the device to malfunc on. 
 

Disinfec on 

Use personal protec ve equipment such as gloves, a smock, and face and eye protec on when handling 
units that are suspected to be contaminated.  

Cau on:  

Disconnect the unit from any power source prior to cleaning the unit. When cleaning the interi

or of the chassis, disconnect the ba ery from the PC Board to prevent damaging the PC Board. 

Part 

 

Cleaning and Disinfec ng

 

Collec on Canister

 

Disposable item, re

use not permi ed. Use new canister for each pa ent.

 

Pa ent Tubing

 

 

 

Disposable item, re

use not permi ed. Use new pa ent tubing for each 

pa ent.

 

HI

D® S ck

 

 

 

Disposable item, re

use not permi ed. Use new HI

D S ck for each pa

ent.

 

Vacuum Pump

 

 

 

Wipe with damp cloth or disinfectant wipe.  Steriliza on not permi ed. 
Vacuum pump should be replaced if contaminated

 

PC Board

 

 

 

Steriliza on not permi ed. PC Board should be replaced if contaminated.

 

Plas c Chassis

 

 

 

Wipe with damp cloth or disinfectant wipe. Steriliza on not permi ed.

 

Clearing The Airway Is Our #1 Priority 

Maintenance 

SSCOR DUET - Model 2314 Series 

Rev W (02/17)  

Summary of Contents for 2314 series

Page 1: ...e Airway Is Our 1 Priority Operating Instructions Maintenance Manual Tel 1 818 504 4054 Tel 1 800 434 5211 INC 11064 Randall Street Sun Valley CA 91352 USA www sscor com shop sscor com Email marketing sscor com techsupport sscor com ...

Page 2: ...all comply with relevant IEC standard e g IEC 60950 for IT equipment and the IEC 60601 series for medical electrical equipment In addi on all such combina ons systems shall comply with the standard IEC 60601 1 1 safety requirements for medical electrical systems Equipment not complying with IEC 60601 shall be kept outside the pa ent environment as defined in the standard Any person who connects ex...

Page 3: ...R Duet is a portable 100V 240V AC constant suc on device with a 12V DC ba ery back up The ba ery is charged by an internal DC dual mode ba ery charger The charger monitors the ba ery charges the ba ery only when necessary shuts down the unit if the ba ery is low and signals ba ery condi on A fully charged ba ery at full capacity will power the unit for 45 minutes 10 SSCOR warrants that each new pr...

Page 4: ...ping Carton Manufacturer Shipping Carton Date of Manufacture Shipping Carton Authorized Representa ve in the European Community Control Panel Power Source Connected Control Panel Pump is On Control Panel Ba ery Level Of Charge Green LED Full Green Yellow LED High Yellow LED Half Yellow Red LED Low Red LED Ba ery Depleted Put device back on charge Control Panel Push On Push Off 4 Clearing The Airway...

Page 5: ...gauge located above the vacuum regulator The vacuum reading should rise to 300mmHg from zero in 3 seconds It should max out at approximately 525mmHg Lower levels of nega ve pressure will be observed at al tude 3 Adjust the vacuum reading to your desired level If the vacuum does not meet or exceed 525mmHg check for a leak in the system i e tubing barb connec ons or loose canister lid Refer internal...

Page 6: ...he desired se ng turn the device off and put it back on charge SSCOR recommends replacement of the ba ery a er 3 years Replacement ba ery may be purchased at shop sscor com PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION Power Source Connected LED not lit Fuse blown Molex connec ons disconnected Power cord disconnected Replace fuse in power supply or input receptacle Reconnect Molex Reconnect cord Does no...

Page 7: ...a ery DC Power Medical Switcher AC Power Rechargeable Sealed Lead Acid SSCOR part 80635 100V 240VAC 47 63Hz Uses 1ea 3A 250V fast ac ng fuse Switch 5 Off Standby On membrane switch AC receptacle 1A Uses 2 2 5A 250V fast ac ng fuses Collec on Canister 13 1200cc ml SSCOR part 48041 Pa ent tubing Vinyl tubing 9 32 ID x 72 L 7 1mm ID x 182 88 cm L SSCOR part 43200 Suc on Tip HI D Big S ck SSCOR part 44...

Page 8: ...essary to open the exhaust door unless fluids have entered the system E PC Board Electrical Circuits Internal Dual Mode Ba ery Charger Do not a empt to service the PC Board Return to SSCOR or authorized service center for service F Power Supply Do not a empt to service the Power Supply Return to SSCOR or authorized service center for service G Func onal ground H AC receptacle Two 250V 2 5A fuses a...

Page 9: ...xhaust filter is soiled remove the two screws holding the filter door in place see page 8 remove the soiled filter and replace with new filter SSCOR part number 90160 Do not reuse any single use disposable parts do not submerge the device into any liquid this will void the warranty and cause the device to malfunc on Disinfec on Use personal protec ve equipment such as gloves a smock and face and e...

Page 10: ...ll be secure at a force equal to or greater than 10G Ver cal Mount Remove the four 20 x 1 1 2 hex head screws from hardware packet Mark the hole placement for a ver cal mount and use a 5 16 drill to drill the holes Coat the screws with Loc te not supplied before securing the Charging Reten on Bracket Put the four 1 1 2 screws through the holes in the charging bracket Model 8323 Rea ach the washers...

Page 11: ...D At the end of the run test turn the device off and reconnect the device to the charging source For a further explana on of the SSCOR ba ery run test visit the SSCOR Website at h p www sscor com training_videos html 3 Determine if the gauge is properly calibrated A er the ba ery has been fully charged when the ba ery condi on LED on the control panel is green connect tubing from the pa ent port of...

Page 12: ... D If the tubing from the device to the canister is clamped to the device 4 remove the tubing from the pa ent port of the canister and take the reading at the distal end of the tubing 5 The airflow at the distal end of the tubing should be at least 48 LPM E If the device s ll does not meet the airflow specifica on call SSCOR technical support 1 800 434 5211 or 1 818 504 4054 to trouble shoot A er ...

Reviews: