Preven ve Care
Observe the following maintenance rou ne to ensure readiness at any me:
1. When the SSCOR aspirator is not in use, keep ba eries on con nuous charge.
2. Test the SSCOR aspirator at regular intervals; See page 6.
3. Make sure the SSCOR aspirator is always clean and ready for use.
4. If the procedure produced an excessive quan ty of
fl
uids, check the vacuum line (7) for evidence of
moisture. If the vacuum line between the pump and canister is moist, it is possible that
fl
uids have reached
the vacuum pump. See Disinfec on Instruc ons below.
5. For technical assistance, call (800) 434
‐
5211 or interna onal +1 818
‐
504
‐
4054. For replacement parts
and accessories, please visit shop.sscor.com.
Note : No part requires lubrica on and lubricants should not be used.
Sanita on
As soon as possible a er use, the single use disposable canister, pa ent tubing and catheter should be
discarded according to local / regional / na onal requirements for the disposal of hazardous waste
materials. Clean using a mild detergent and if necessary disinfect with a mild disinfectant such as 10 to 1
bleach and rinse using clear water to remove any residue.
If the exhaust
fi
lter is soiled, remove the two screws holding the
fi
lter door in place (see page 8), remove
the soiled
fi
lter and replace with new
fi
lter (SSCOR part number 90160).
Do not reuse any single use disposable parts; do not submerge the device into any liquid, this will void the
warranty and cause the device to malfunc on.
Disinfec on
Use personal protec ve equipment such as gloves, a smock, and face and eye protec on when handling
units that are suspected to be contaminated.
Cau on:
Disconnect the unit from any power source prior to cleaning the unit. When cleaning the interi
‐
or of the chassis, disconnect the ba ery from the PC Board to prevent damaging the PC Board.
Part
Cleaning and Disinfec ng
Collec on Canister
Disposable item, re
‐
use not permi ed. Use new canister for each pa ent.
Pa ent Tubing
Disposable item, re
‐
use not permi ed. Use new pa ent tubing for each
pa ent.
HI
‐
D® S ck
Disposable item, re
‐
use not permi ed. Use new HI
‐
D S ck for each pa
‐
ent.
Vacuum Pump
Wipe with damp cloth or disinfectant wipe. Steriliza on not permi ed.
Vacuum pump should be replaced if contaminated
PC Board
Steriliza on not permi ed. PC Board should be replaced if contaminated.
Plas c Chassis
Wipe with damp cloth or disinfectant wipe. Steriliza on not permi ed.
9
Clearing The Airway Is Our #1 Priority
Maintenance
SSCOR DUET - Model 2314 Series
Rev W (02/17)