WARNING: Do not a empt to service this equipment if you are not a quali
fi
ed medical repair technician
To open or close the unit,
fi
rst remove the canister holder:
Remove the set screws on the canister bracket.
Li the canister holder up and out of the main chassis.
The clamshell design enables access to internal components.
All 6
‐
32 Phillips head fastening screws are located in back of
the unit.
REMOVE MAIN CHASSIS SCREWS
(A) 1 screw under the canister holder
(B) 2 screws in the back of the handle
(B) 3 screws in the back of the main chassis.
(C) It should not be necessary to open the exhaust door
unless
fl
uids have entered the system
(E) PC Board
‐
Electrical Circuits (Internal Dual Mode Ba ery Charger). Do not a empt to service the PC Board.
Return to SSCOR or authorized service center for service.
(F) Power Supply. Do not a empt to service the Power Supply. Return to SSCOR or authorized service center
for service.
(G) Func onal ground.
(H) AC receptacle. Two 250V
‐
2.5A fuses are in the receptacle fuse drawer.
(I) Ba ery Bracket
(J) Ba ery, 12V, Sealed Lead Acid
(K) Wiring connec on to the control panel
(L) Vacuum pump. Return to SSCOR or authorized service center for service. Do not a empt to service the
vacuum pump.
(M) Exhaust barb
(N) Vacuum barb