Remote Programming
96
CTC100 Programmable Temperature Controller
while (Out1.Tune.Mode != Off) { pause 1 s }
Sy stem setup
system.com.baud { 2400, 4800, 9600, 14400, 19200, 28800, 38400, 57600,
115200, 250000 }
Sets the baud rate for the RS-232 and USB-to-PC interfaces. The interface always has no parity,
8 bits, and 1 stop bit.
system.com.flow ctl { None, RTS/CTS }
Enables or disables hardware flow control for the RS-232 and USB-to-PC interfaces.
system.com.verbose { Low, Medium, High }
Determines how the CTC100 replies when a remote instruction is processed.
Low
: the CTC100 only replies when a query is processed.
Medium
: messages are also produced whenever an error occurs.
High:
messages are also sent whenever an instruction sets a parameter, and the messages
include the names of parameters that are set or queried.
V erbose level
2A ?
xyz
2A = 37.47
Low
37.4722
(none)
(none)
Medium
37.4722
uction
(none)
High
2A.Value = 37.4722
2A.Value = 37.47
system.display.Bright { Min, 2, 3, 4, 5, 6, Max }
Sets the brightness of the front-panel display backlight.
system.display.Figures { 0, 1, 2, 3, 4, 5, 6 }
Sets the maximum number of figures that appear after the decimal point in the replies to remote
queries of floating-point channel values, as well as on the front panel display. This setting does not
affect how data is stored internally.
Each channel also has its own Figures setting that can override this global value.
system.display.Monitors { Show, Hide, Off }
ed by
the getOutput command. These channels display printed circuit board (PCB) temperatures for the
I/O cards as well as heater current, voltage, and resistance. If set to Hide, the monitor channels
are not displayed on-screen but are still logged and still appear in the response to getOutput. If set
to Off, the c
in the response to getOutput.
system.display.Stats { Off, On }
Controls whether statistics are visible in the plot. If set to On and the plot type is single or
multiple, the average and standard deviation for each channel for which statistics collection has
been enabled (with
channel.
stats
) is shown next to the channel name. Ponytail plots do not
show the average and standard deviation, but instead show the offset of each channel, if stats
display has been enabled.
Summary of Contents for CTC100
Page 1: ...Version 2 1 May 14 2019 User Manual CTC100 Cryogenic Temperature Controller...
Page 6: ......
Page 8: ......
Page 12: ......
Page 25: ...Operation 13 CTC100 Programmable Temperature Controller...
Page 85: ...Operation 73 CTC100 Programmable Temperature Controller To disable cascade cont select it...
Page 128: ......
Page 129: ...Remote Programming 117 CTC100 Programmable Temperature Controller...
Page 130: ......
Page 142: ......
Page 150: ......
Page 176: ......