SR Instruments SR Scales SR7005i Operating And Service Manual Download Page 9

Model SR7005

i

 Wall Mount In-Room Mini Scale 

Operating and Service Manual 
Part No. MAN7005i_200520 

Page 9 of 16 

 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel:  716-693-5977     Fax:  716-693-5854    URL:  

www.srscales.com

 

email:  

[email protected]

       Copyright 2020

 

S

Instruments, Inc. 

BUTTON FUNCTIONS 

 

 
 

WEIGH 

Press and hold to zero scale. Button is used to zero the system before placing the patient on the 
scale. Ensure that nothing is in contact with the weighing surface during this procedure. 

 

SEND (PRINTER) 

Press to send stored values to printer. Output values include time, date, and weight. If BMI was 
calculated, BMI and height will be included in the output values. 
 
 

RECALL 

Press to recall the last stored weight. The stored weight will be erased when the scale is zeroed 
or another stable weight is stored. 
 
 

BMI 

Press to calculate BMI. When the “

BMI

” is pressed, the default value “

HT

 

= 65 in

” or “

HT = 

165  cm

”  is  displayed.  If  there  is  no  stored  stable  weight,  the  display  will  indicate  “

NO 

WEIGHT DATA

” and then go back to the weigh screen “

WT = 0.0 Lb

”. 

 

MENU 

Press Menu to toggle through the menu options.

 

Setting

 

the

  UNITS

:  Use 

UP

  or 

DOWN

  arrow  buttons  to  select  “

Lb/  in

”  or  “

Kg/  cm

”.  Press 

ENTER

 to save changes. 

Setting 

ON  TIME

:  Use 

UP

  or 

DOWN

  arrow  buttons  to  adjust  the  “

ON  TIME

”.  The  “

ON 

TIME

”  may  be  set  from  30  to  300  seconds  in  30  second  increments.  Press 

ENTER

  to  save 

changes. 

Setting

  TIME 

and 

DATE

:  Use  the 

UP

  arrow  button  to  select  digit.  To  change  digit  use  the 

DOWN

 arrow button. Press 

ENTER

 to save changes. 

NOTE

: When selected, the year position defaults to “00” 

Figure 3:  Button Display 

Continued next page 

Summary of Contents for SR Scales SR7005i

Page 1: ...00520 Page 1 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S Instru...

Page 2: ...F CONTENTS PACKING CHECKLIST 3 INSTALLATION 3 MICROBAN ANTIMICROBIAL PROTECTION 5 SYSTEM DESCRIPTION AND INTENDED USE 6 CLEANING AND DISINFECTING 7 STORAGE 7 SPECIFICATIONS 8 BUTTON FUNCTIONS 9 BASIC...

Page 3: ...tends 23 into the room hallway when folded down Hallway must be wide enough to allow suitable access and egress for personnel using the hallway If using the AC power supply a wall outlet must be locat...

Page 4: ...walls use a stud finder and locate any studs in the mounting area STEP b Measure up from the floor 27 5 8 and make a mark at a stud location This will be the location of bottom edge of the wall mount...

Page 5: ...ons STEP b For stud wall mounting locations without a stud use a screwdriver to drive the provided self drilling wall anchors into the wall It is not recommended to use a power drill to set the wall a...

Page 6: ...t s weight is displayed on a 16 character dot matrix LCD w LED backlight The weight data may be viewed either in pounds or kilograms with a displayed resolution of 0 1 for each The SR7005i Wall Mount...

Page 7: ...sodium hypochlorite or 70 isopropyl alcohol are suitable for display user interface and other scale contact surfaces After disinfecting use a soft cloth dampened with clean water and dry with a soft...

Page 8: ...AUTO POWER DOWN Adjustable between 30 to 300 seconds AVERAGING Automatic digital filter POWER SUPPLY Six 6 C cell batteries Replace battery indicator on Display Medical grade AC power supply CALIBRAT...

Page 9: ...last stored weight The stored weight will be erased when the scale is zeroed or another stable weight is stored BMI Press to calculate BMI When the BMI is pressed the default value HT 65 in or HT 165...

Page 10: ...ng time and date BASIC SYSTEM OPERATION SETTING SYSTEM ZERO Scale will auto zero when the platform is first lowered Make sure the scale is free and clear of any obstructions When platform is left down...

Page 11: ...lied mass on the system into minute changes in electrical resistance These changes in resistance produce a voltage difference across the Wheatstone bridge which is amplified by the operational amplifi...

Page 12: ...NIST NOTE DO NOT USE barbells or uncalibrated weights STEP 2 Zero the scale by pressing and holding the ZERO button STEP 3 Place the calibrated weights on the scale Wait for scale to stabilize note sc...

Page 13: ...button until HALF 250 0 lb is displayed Set the HALF value to the actual quantity of calibrated weight being used for Half Scale Use the UP arrow button to select the digit and the DOWN arrow button t...

Page 14: ...attery Wait five seconds then re install battery and try the WEIGH button again The display lights appear to work but do not respond to button activation Check to ensure the membrane switch label is c...

Page 15: ...liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In n...

Page 16: ...al Part No MAN7005i_200520 Page 16 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstrum...

Reviews: