INSTALLATION (Cont’d)

 

 

STEP (d):

 Mount the wall bracket to the wall by using the provided #8 x 1-1/2” Phillips flat head screws 

into the self-drilling wall anchors or using customer supplied appropriate flat head screws long enough to 
anchor into a wall stud or the concrete anchors. 

STEP (e):

 After the wall bracket is securely fastened to the wall, carefully lift the SR7005i scale and place 

it onto the wall mounting bracket. Proceed to Step 4. 

 
 
 
 
 
 
 
 

USING THE CROSSBAR MOUNTING HOLES 

STEP (a):

 Place the scale against the wall and carefully lower the platform. Make a mark at a stud location 

for the two holes 16” apart (Figure 2). 

For concrete walls, follow the same procedure, marking the two hole locations. 

STEP (b):

 For stud wall mounting locations without a stud, use a screwdriver to drive the provided self-

drilling wall anchors into the wall. It is not recommended to use a power drill to set the wall anchors. 

For concrete wall mounting, install appropriate wall anchors at all marked screw locations. 

STEP (c):

 Attach the crossbar to the wall by using the provided #8 x 1-1/2” Phillips flat head screws into 

the  self-drilling  wall  anchors  or  using  customer  supplied  appropriate  flat  head  screws  long  enough  to 
anchor into a wall studs or the concrete anchors. 

STEP 4:

 Plug in the AC power supply and/or remove the battery cover and remove the plastic battery strip 

to activate the six (6) “C” cell batteries. Reattach the battery cover. 

 
 
 

MICROBAN ANTIMICROBIAL PROTECTION 

 
Microban® antimicrobial technology has been added to help reduce the growth of bacteria as part of a 
medical center's approach to creating a cleaner healthcare environment. For more information, visit 

https://www.microban.com/sr-scales 

 
 

IMPORTANT 

THE WALL BRACKET IS INTENDED TO KEEP THE SCALE 

FROM PULLING OFF OF THE WALL. ENSURE THAT THE 

ENTIRE WEIGHT OF THE SCALE IS ON THE FLOOR AND 

NOT ON THE WALL BRACKET. 

Summary of Contents for SR Scales SR7005i

Page 1: ...00520 Page 1 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S Instru...

Page 2: ...F CONTENTS PACKING CHECKLIST 3 INSTALLATION 3 MICROBAN ANTIMICROBIAL PROTECTION 5 SYSTEM DESCRIPTION AND INTENDED USE 6 CLEANING AND DISINFECTING 7 STORAGE 7 SPECIFICATIONS 8 BUTTON FUNCTIONS 9 BASIC...

Page 3: ...tends 23 into the room hallway when folded down Hallway must be wide enough to allow suitable access and egress for personnel using the hallway If using the AC power supply a wall outlet must be locat...

Page 4: ...walls use a stud finder and locate any studs in the mounting area STEP b Measure up from the floor 27 5 8 and make a mark at a stud location This will be the location of bottom edge of the wall mount...

Page 5: ...ons STEP b For stud wall mounting locations without a stud use a screwdriver to drive the provided self drilling wall anchors into the wall It is not recommended to use a power drill to set the wall a...

Page 6: ...t s weight is displayed on a 16 character dot matrix LCD w LED backlight The weight data may be viewed either in pounds or kilograms with a displayed resolution of 0 1 for each The SR7005i Wall Mount...

Page 7: ...sodium hypochlorite or 70 isopropyl alcohol are suitable for display user interface and other scale contact surfaces After disinfecting use a soft cloth dampened with clean water and dry with a soft...

Page 8: ...AUTO POWER DOWN Adjustable between 30 to 300 seconds AVERAGING Automatic digital filter POWER SUPPLY Six 6 C cell batteries Replace battery indicator on Display Medical grade AC power supply CALIBRAT...

Page 9: ...last stored weight The stored weight will be erased when the scale is zeroed or another stable weight is stored BMI Press to calculate BMI When the BMI is pressed the default value HT 65 in or HT 165...

Page 10: ...ng time and date BASIC SYSTEM OPERATION SETTING SYSTEM ZERO Scale will auto zero when the platform is first lowered Make sure the scale is free and clear of any obstructions When platform is left down...

Page 11: ...lied mass on the system into minute changes in electrical resistance These changes in resistance produce a voltage difference across the Wheatstone bridge which is amplified by the operational amplifi...

Page 12: ...NIST NOTE DO NOT USE barbells or uncalibrated weights STEP 2 Zero the scale by pressing and holding the ZERO button STEP 3 Place the calibrated weights on the scale Wait for scale to stabilize note sc...

Page 13: ...button until HALF 250 0 lb is displayed Set the HALF value to the actual quantity of calibrated weight being used for Half Scale Use the UP arrow button to select the digit and the DOWN arrow button t...

Page 14: ...attery Wait five seconds then re install battery and try the WEIGH button again The display lights appear to work but do not respond to button activation Check to ensure the membrane switch label is c...

Page 15: ...liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In n...

Page 16: ...al Part No MAN7005i_200520 Page 16 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstrum...

Reviews: