STEP 1:

 Test the battery by placing the power switch to the “ON” 

position. The light indicates good battery. If the light flickers or 
does not come on, replace the batteries as described in the battery 
procedure. Place the power switch to the “OFF” position after the 
battery condition has been checked.

STEP 2:

 To test DC holiday detector connect the high voltage (red 

cable) test lead to the DC input terminal on the top left of the JeepMeter. 

To test pulse holiday detector connect the high voltage (red cable) test 

lead to the pulse input terminal on the top center of the unit.

STEP 3:

 Connect the other end (alligator clip) of the (red wire) high 

voltage test lead to a holiday detector electrode or the metal portion of 
the electrode wand. 

STEP 4:

 Connect the ground (black cable) test lead to the ground input 

terminal located on the top right of the JeepMeter.

STEP 5:

 Connect the other end (alligator clip) of the ground (black cable) 

test lead to the bare metal of the pipe if possible, otherwise to the bare 
wire of the holiday detector ground cable. (Connecting to the bare pipe 
gives a more accurate reading of the actual working voltage impressed 
on the coating).

CAUTION!

 TO PREVENT DAMAGE TO THE JEEPMETER, THE 

INPUT PULSE / DC SWITCH MUST BE PROPERLY POSITIONED.

Summary of Contents for JeepMeter

Page 1: ...Spy Jeepmeter operating instructions PIPELINE INSPECTION COMPANY picltd com...

Page 2: ...work which supplies buildings used for domestic purpose is a registered trademark of Pipeline Inspection Company Houston Texas 77055 United States A copy of this Operating Instructions Manual can be d...

Page 3: ...ration 4 Principles of Operation 5 JeepMeter Overview 6 Battery Instructions 8 Controls Testing Instructions 9 JeepMeter Operation 10 Troubleshooting Guide 12 Technical Specifications 13 Maintenance R...

Page 4: ...usage Recalibration of our equipment should be performed ONLY at a SPY Authorized Service Center due to the unique calibration process test equipment utilized and properly trained SPY technicians We...

Page 5: ...JM enables the operator to measure DC up to 7 000 kV or pulse up to 35 kV Holiday detectors output both types of waveforms depending on the type of coating they are being used on For example the SPY...

Page 6: ...er provides the ultimate in accuracy and versatility It features a digital display of measured voltages The precision measurement displayed eliminates interpretation errors and is perfect for every da...

Page 7: ...Input Terminal 3 High Voltage Pulse Input Terminal 4 Quick Change Battery Compartments 5 Ground Input Terminal 6 Input Pulse DC Bat Switch 7 Output Voltage LCD Display Battery Indicator 8 Extendable F...

Page 8: ...ing the notch on the bottom of the battery compartment lift up and pull out to slide the battery trays out STEP 3 Insert two new 9 volt batteries into the battery trays and push the battery tray firml...

Page 9: ...operating instructions controls POWER On Off INPUT Pulse DC Kilovolts Pipeline Inspection Co testing instructions Test the strength of the batteries by placing the power switch to the ON position The...

Page 10: ...test lead to the pulse input terminal on the top center of the unit STEP 3 Connect the other end alligator clip of the red wire high voltage test lead to a holiday detector electrode or the metal por...

Page 11: ...6 Set the input switch to Pulse or DC position to match which input terminal is being used STEP 7 After and ONLY AFTER double checking that all the connections in Steps 1 through 6 are correct place...

Page 12: ...elected on the Pulse DC switch The 780 is DC The 785 and 790 are Pulse Wrong Connection High Voltage Pulse and DC Voltages have different circuitry to measure voltages and therefore must have differen...

Page 13: ...e Voltage Range 0 35 kV Pulse and 0 5 kV DC Measurement Accuracy 5 75 deg F Display Type LCD Battery Two 9 Volt Batteries Unit Weight w battery only 2 6 Unit Dimensions L x W x H 8 25 x 8 75 x 4 25 Fa...

Page 14: ...ned by anyone other than a SPY Authorized Service Center or SPY manufacturing facility In the event the unit fails to operate properly please return your unit to a SPY Authorized Service Center or SPY...

Page 15: ...for examination to establish buyer s claim Except with SPY s prior written approval SPY shall not be liable a for the cost of repairs alterations or replacements or any expense connected therewith ma...

Page 16: ...d com related equipment SPY MODEL PJM POCKET JEEPMETER SPY MODEL DCPJM DC POCKET JEEPMETER Please call our office at 713 681 5837 or email us at sales picltd com for more product details or to place a...

Page 17: ...picltd com 17 User notes...

Page 18: ...18 picltd com User notes...

Page 19: ...picltd com 19...

Page 20: ...PIPELINE INSPECTION COMPANY picltd com...

Reviews: