Clearing Auto-Stored Data

1.  Press and hold 

POWER

 to turn the meter 

OFF

.

2.  Press and hold 

REC

. While still holding 

REC

, press and release 

POWER

. “CLr” appears on the LCD. The LCD displays “SURE 5, 

4, 3, 2, 1, 0.” (The meter allows 5 seconds to abort the operation 

by releasing 

REC

.) Continue holding 

REC

 until the meter returns to 

Normal Mode. The memory is now cleared. 

Manually Storing Data

1.  Press 

MEM

 to store up to 99 data point using manual storage. 

Reading Manually-Stored Data

1.  Press 

RECALL

. “READ” appears on the LCD. 

2.  Press ▲ or ▼ to select the data set. The data will display on the LCD. 

If there are no data in a data set, “MEM00” appears on the LCD. 

Press 

RECALL

 to exit Recall Mode. 

Clearing Manually-Stored Data

1.  Press and hold 

RECALL

. While still holding 

RECALL

, press and 

release 

POWER

. “CLr” appears on the LCD. The LCD displays “SURE 

5, 4, 3, 2, 1, 0.” (The meter allows 5 seconds to abort the operation by 

releasing 

RECALL

.) Continue holding 

RECALL

 until the meter returns 

to Normal Mode. The memory is now cleared. 

BATTERY REPLACEMENT

1.  Remove the screw at the top of the battery cover and lift the cover off 

the meter.

2. Insert 4 AA batteries, ensuring correct polarity.
3. Replace the battery cover and tighten the screw to secure. 

Note… 

This meter can use AA rechargeable batteries. A separate charger 

must be provided for rechargeable batteries because the meter is not 

designed to recharge batteries. 

MAINTENANCE

The following actions will keep the meter in good working condition and 

ensure correct measurement technique: 
•  Wind blowing across the microphone may cause additional noise and 

affect the reading. When taking measurements in a windy environment, 

the windscreen must be used to prevent interference.

Summary of Contents for 850013

Page 1: ...Advanced Datalogging Sound Meter 850013 Instruction Manual...

Page 2: ...other means that do not yet exist or may be developed including photocopying recording or any information storage and retrieval system without the express permission from Sper Scientific INTRODUCTION...

Page 3: ...measurements in science and engineering the decibel is most commonly recognized as a measure of the intensity of a sound wave The decibel is a unit of power ratio with the number of units equal to a...

Page 4: ...battery power is almost empty the empty battery symbol will disappear from the LCD Press MEM or REC Lo Bat appears on the LCD If the meter is in recording mode it will stop recording For battery repl...

Page 5: ...utton 13 Battery compartment on back 14 A C Sets frequency rating to A or C scale 15 CAL potentiometer Calibration control for level calibration adjustment 16 External DC 9V power supply jack Connect...

Page 6: ...ttery fluid leakage After reinstalling the batteries set the date and time CALIBRATION The meter has been factory calibrated Annual calibration is recommended for best results 1 Prepare for calibratio...

Page 7: ...N MAX again to view the minimum 4 To exit this mode and clear all maximum and minimum values press and hold MIN MAX until MAX MIN disappears from the display Note Maximum and minimum values cannot be...

Page 8: ...RECALL press and release POWER CLr appears on the LCD The LCD displays SURE 5 4 3 2 1 0 The meter allows 5 seconds to abort the operation by releasing RECALL Continue holding RECALL until the meter r...

Page 9: ...her applications Close other applications If the problem is still not resolved restart the computer and run the software again 2 Saving the graph to a file that can be used in Excel When saving the gr...

Page 10: ...nstalling corresponds to the Operating System platform running on your machine 3 Aside from the general hardware requirements some computer programs also have software requirements prior to installati...

Page 11: ...n completed and the setup wizard has terminated reboot your machine before launching the newly installed software Main Menu File_Open Retrieves files from the disk File_Save Saves the active window th...

Page 12: ...en selected the mouse cursor will become a cross sign when moving through the graph Click on the graph to mark a cross sign on the graph When selected the mouse cursor will become a I sign text tool w...

Page 13: ...al time graph is not allowed to open the popup menu To zoom in on the graph using the mouse Click the left mouse button and drag the cursor to select the new graph boundaries Release the mouse button...

Page 14: ...icon from the tool bar 3 The save dialog window appears on the screen Select the file name and file type to save Note There are three file type options binary ghf text txt and Excel format csv Of the...

Page 15: ...OVER indicates that the input is over the upper range limit UNDER indicates that the input is less than the lower range limit Datalogging capacity 31 000 records Operating temperature 0 to 40 C 32 to...

Page 16: ...y does not cover probes batteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will break the waterproof seal...

Reviews: