OPERATING INSTRUCTIONS
Sensor Types
The meter works with optional pressure sensors.
When changing the sensor, re-calibration is not required.
The optional pressure sensor are:
• 2 Bar pressure sensor
• 5 Bar pressure sensor
• 10 Bar pressure sensor
• 20 Bar pressure sensor
• 50 Bar pressure sensor
• 100 Bar pressure sensor
• 200 Bar pressure sensor
Changing the Sensor
1. Turn the meter off.
2. Push and hold the following 3 buttons at the same time:
HOLD
Button,
MAX/MIN
Button,
SENSOR TYPE
Button
3. Continue to hold the 3 buttons, while pushing the
POWER
Button. Release the buttons and the display will indicate the
previous sensor type.
4. Push the
SENSOR TYPE
Button to select the sensor (2, 5, 10,
20, 50, 100, or 200 Bar). When the display indicates the desired
sensor, push the Zero button. The new sensor type will be
stored in memory, even after the unit has been turned off.
5. Turn off the meter. This completes the sensor selection.
Measuring Procedures
1. Connect the Pressure Sensor Plug to the meter’s Sensor
Input Socket.
2. Press the
POWER
Button.
3. Press the
SENSOR TYPE
Button to check that the meter’s
sensor type is correct.
4. Press the
UNIT
Button to select the desired unit of measure:
Bar, Psi, Kg/cm
2
, mm/Hg, inch/Hg, meter/H
2
0, inch/H
2
0, or
Atmosphere.
5. Press the
ZERO
Button and the display reading will show zero.
6. Attach the Pressure Sensor Port Connector to the object to
be measured.
7. Apply pressure and the meter will indicate the pressure value.