background image

4

5

OPERATING INSTRUCTIONS
Sensor Types

The meter works with optional pressure sensors. 

 

When changing the sensor, re-calibration is not required.

The optional pressure sensor are:
• 2 Bar pressure sensor
• 5 Bar pressure sensor
• 10 Bar pressure sensor
• 20 Bar pressure sensor

• 50 Bar pressure sensor
• 100 Bar pressure sensor
• 200 Bar pressure sensor

Changing the Sensor

1.  Turn the meter off.
2.   Push and hold the following 3 buttons at the same time: 

 

HOLD

 Button, 

MAX/MIN

 Button, 

SENSOR TYPE

 Button

3.  Continue to hold the 3 buttons, while pushing the 

POWER

 

Button. Release the buttons and the display will indicate the 

previous sensor type.

4.  Push the 

SENSOR TYPE

 Button to select the sensor (2, 5, 10, 

20, 50, 100, or 200 Bar). When the display indicates the desired 

sensor, push the Zero button. The new sensor type will be 

stored in memory, even after the unit has been turned off.

5.  Turn off the meter. This completes the sensor selection.

Measuring Procedures

1.  Connect the Pressure Sensor Plug to the meter’s Sensor 

 

Input Socket.

2.  Press the 

POWER

 Button.

3.  Press the 

SENSOR TYPE

 Button to check that the meter’s 

sensor type is correct.

4.  Press the 

UNIT

 Button to select the desired unit of measure: 

Bar, Psi, Kg/cm

2

, mm/Hg, inch/Hg, meter/H

2

0, inch/H

2

0, or 

Atmosphere.

5.  Press the 

ZERO

 Button and the display reading will show zero.

6.  Attach the Pressure Sensor Port Connector to the object to 

 

be measured.

7.  Apply pressure and the meter will indicate the pressure value.

Summary of Contents for 840065

Page 1: ...Wide Range Pressure Meter 840065 Instruction Manual...

Page 2: ...do not yet exist or may be developed including photocopying recording or any information storage and retrieval system without the express permission from Sper Scientific INTRODUCTION 3 PANEL DESCRIPTI...

Page 3: ...ulic servo controls refrigeration air conditioning and food processing Features eight units of measure Bar Psi Kg cm mm 1meter H20 inch H20 Atmosphere min max hold and auto power off PANEL DESCRIPTION...

Page 4: ...ype 4 Push the SENSOR TYPE Button to select the sensor 2 5 10 20 50 100 or 200 Bar When the display indicates the desired sensor push the Zero button The new sensor type will be stored in memory even...

Page 5: ...nt reading Measuring Procedures Overview 1 Connect the Sensor Plug to the meter s Input Socket 2 Turn on the meter 3 Select the display unit 4 Zero the meter by pressing the ZERO Button 5 Connect the...

Page 6: ...it to warm up for two minutes 2 Push the ZERO Button 3 Connect the pressure regulator to a nitrogen bottle and the reference gauge to the pressure regulator 4 Press and hold both the MAX MIN Button a...

Page 7: ...ATIONS TRANSDUCERS SPECIFICATIONS METER Circuit Microprocessor LSI circuit Sensor type Works with optional 2 5 10 20 50 100 200 bar sensor new calibration is not necessary when the sensor is changed M...

Page 8: ...does not cover probes batteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will break the waterproof seal a...

Reviews: