8
Temperature and Humidity Offset
The meter contains precision temperature and humidity
sensors which can be adjusted to ensure precise accuracy.
They can be adjusted within ± 9.9°C, ± 9.9°F, and 9.9% RH.
1. If the meter is on, press and hold
PWR
to turn the meter
OFF
.
2. Press and hold down
MODE
and
BKLT
with your left hand.
3. While continuing to hold
MODE
and
BKLT
press and hold
down
PWR
and
MAX/MIN
with your right hand. “0.0” and
“SEt” display.
4. Press
HOLD
to increase the value.
5. Press
BKLT
to decrease the value.
6. Press
UNIT
to toggle between °C and °F.
7. Press
SET
to adjust the RH. “0.0%” displays.
8. Press
HOLD
to increase the value.
9. Press
BKLT
to decrease the value.
10. Press and hold
SET
to save the values. “SAVE” displays
and the values are stored.
11. Press and hold
PWR
to turn the meter
OFF
.
Note…
When pressing
HOLD
or
BKLT
to adjust a value, you can
scroll more quickly by holding down the button.